o p e r a t i o n
4
S y s t e m i n t e r f a c e s
PHILIPS VUELINK SETUP
1. Select the Philips VueLink selection from the Serial Output menu on the Rad-87. After
selecting, choose the preferred settings by stepping through menu options.
2.
Connect one end of the VueLink cable to the Serial Output connector on the back of the
Rad-87.
3.
Connect the other end of the VueLink cable to the VueLink module and insert the module
into the Philips/Agilent monitor rack.
4.
The SpO
2
and pulse rate values will automatically appear on the Philips/Agilent monitor.
5.
In order for the pleth waveform to be displayed on the Philips/Agilent monitor and for the
Philips/Agilent monitor to indicate the alarm conditions measured by the pulse oxime-
ter, the user must configure the Philips/Agilent monitor. Please see the Philips/Agilent
Operator’s manual for complete instructions.
6.
The Rad-87 Pulse CO-Oximeter can be set up to audibly indicate all patient alarms while
communicating with the Philips/ VueLink module. Use the Interface Alarms setting in the
Output menu to enable and disable audible alarms on the Rad-87
.
RADNET SETUP
1.
Connect one end of the serial cable to the Serial Output connector on the back of the
Rad-87.
2.
Connect the other end of the serial cable to the RadNet Interface Module connector
3.
Turn the RadNet Interface Module on.
4.
Select the ASCII 2 selection from the Serial options on the Rad-87.
5.
A proper connection is shown by the RadNet Interface Module's Online LED being
solid.
6.
With a properly configured RadNet Interface Module, the Rad-87 will automatically
display the SpO2, PVI and Pulse Rate parameters on the screen at the RadNet Central
Station.
7.
The Rad-87 Pulse CO-Oximeter can be set up to audibly indicate all patient alarms
while communicating with the RadNet Interface module.
PATIENT SAFETYNET SETUP
1.
Select the ASCII 2 selection from the Serial options on the Rad-87.
2.
Contact Masimo installation personnel for proper installation guidance.