Rad-87 Pulse CO-Oximeter Operator’s Manual
5-3
5-3
5
a l a r m s a n d m e s s a g e s
ALARM LIMIT: USER CONFIGURABLE SETTINGS (CONTINUED)
SETTING
RANGE
PI High Limit
The PI high alarm limit can be set anywhere between 0.04% and 19%,
then “----” (Off). Adjust in increments of 0.01% step size between 0.04%
and 0.1%, 0.1% step size between 0.1% and 1% and 1% step size
between 1% and 19%. When PI is placed in the “----” (Off) setting, the PI
High Limit alarm is disabled.
PI Low Limit
*
The PI low alarm limit can be set anywhere between "---" (Off), then
0.03% and 18%. Adjust in increments of 0.01% step size between
0.03% and 0.1%, 0.1% step size between 0.1% and 1% and a 1% step
size between 1% and 18%. When PI is placed in the “----” (Off) setting,
the PI Low Limit alarm is disabled.
PVI High Limit
The PVI high alarm limit can be set anywhere between 2% and 99%,
then “----” (Off). Adjust in increments of 1% step size. When PVI is
placed in the “----” (Off) setting, the PVI High Limit alarm is disabled.
PVI Low Limit
*
The PVI low alarm limit can be set anywhere between "---" (Off), then 1
and 98%. Adjust in increments of 1 step size. When PVI is placed in the
“----” (Off) setting, the PVI Low Limit alarm is disabled.
Pulse Rate High
Limit (BPM)
The pulse rate high alarm limit can be set anywhere between
35 BPM and 235 BPM, with a 5 BPM step size.
Pulse Rate Low
Limit (BPM)
*
The pulse rate low alarm limit can be set anywhere between
30 BPM and 230 BPM, with a 5 BPM step size.
*
The low alarm limit must always be set below the high alarm setting. Attempting to
set the high alarm limit below the low alarm limit, the low alarm limit will automatically
adjust the low limit to the next setting below the newly entered high alarm limit setting.
NOTE:
Pressing and holding down the up and down buttons allow for the rapid scrolling
of changing alarm limits.
NOTE:
If there is a loss of power for any length of time, the Alarm settings will be set
back to the User set defaults. If the user has not utilized this option, then all
parameters/measurements will be set back to the factory defaults.