Rad-87 Pulse CO-Oximeter Operator’s Manual
4
o p e r a t i o n
PARAMETER/MEASUREMENT SELECTION
The bottom fi eld of any display screen may be confi gured to show PVI (see Parameter
/Measurement Select menus). Once PVI is confi gured to show on a screen different
from the default location, the alarm limit menu for PVI can be accessed from the new
screen by pressing the Alarm Limits Button.
PVI may be confi gured to show on one, two or three screens simultaneously (see Param-
eter/Measurement Select menus). However, when placing PVI on multiple screens, the
alarm limit menu for PVI must be accessed from the default location.
NOTE:
When one parameter/measurement is confi gured for all three screens, the other
parameter/measurement will not show or alarm. A parameter/measurement must
show on a screen so that the alarm limit options can be accessed and set for that
parameter/measurement.
NOTE:
SpMet and/or SpCO must be confi gured and PVI set to show on Screen 3 for
Screen 3 to show. If SpMet and SpCO are not confi gured and PVI is not set
to show on Screen 3, Screen 3 will not show. The display screens only move
between Screen 1 and Screen 2.
Display Screens Showing PVI on Screen 1, 2 and 3
Screen
1
Home display screen
Screen
2
Screen
3
6
3
13
%SpCO
1.60
3
3