Rad-87 Pulse CO-Oximeter Operator’s Manual
4-31
4
o p e r a t i o n
SETUP MENU LEVEL
3
(CONTINUED)
SERIAL OUTPUT
BUTTONS
SETTING
Press the Enter
Button again to
move to the next
menu.
12x
Default
AS2 (ASCII 2)
Use Up or Down Button to move
between settings
AND
press the Enter Button to accept
the setting and move to the next
menu.
OR
press the Display Button to exit
without saving the new setting
and to return to the home display
screen.
AS1 (ASCII 1)
PHL
(Philips VueLink)
INTERFACE ALARMS
When Rad-87 is interfaced to another system and the Interface Alarms are set to "Off",
SpO2 and BPM audible alarms are muted at the Rad-87 and active at the interfaced
system. This prevents both systems from producing SpO2 and BPM audible alarms at the
same time.
NOTE:
The Rad-87 reverts to Interface Alarms "On" during power interruptions or when the
interface connection is lost. This ensures that the Rad-87 provides SpO2 and BPM
audible alarms when connection to the interfaced system becomes compromised.
BUTTONS
SETTING
Press the Enter
Button again to
move to the next
menu.
13x
Default
On
NOTE:
LCD shows
SpO2 and BPM alarms
On.
Use Up or Down Button to
move between settings
AND
press the Enter Button to
accept the setting and
move to the next menu.
OR
press the Display Button to
exit without saving the new
setting and to return to the
home display screen.
Off (only SpO2/BPM)
NOTE:
LCD shows
SpO2 and BPM alarms
Off.