Rad-87 Pulse CO-Oximeter Operator’s Manual
7-5
SERIAL PRINTER SETUP
To print the SpO
2
and pulse rate data in ASCII 2 format on a serial printer, simply connect
the serial printer to the serial port and set output mode to ASCII 2. Once serial communi-
cation is established, the Rad-87 will automatically start printing the ASCII 2 text data.
WARNING:
ALL EXTERNAL DEVICE CONNECTIONS TO THE RS-232 SERIAL
PORT MUST BE IEC-60950 COMPLIANT.
Nurse call specifications
The nurse call features are accessible via the 1/4” round female connector on the back
of the device.
NURSE CALL
The nurse call feature on the Rad-87 Pulse CO-Oximeter is based on the relay closing
or opening depending on alarm, Low Signal IQ events or both. In addition the nurse call
polarity can be inverted to accommodate various nurse call stations requirements.
The nurse call relays have the following electrical specification per switch:
PARAMETER
SPECIFICATION
MAX VOLTAGE
36 VDC or 24 VAC peak
WARNING:
THE NURSE CALL FEATURE IS DISABLED WHEN THE AUDIBLE
ALARMS ARE SILENCED AND NURSE CALL SETTING IS SET TO
“ALARMS”.
7
s p e c i f i c a t i o n s