Rad-87 Pulse CO-Oximeter Operator’s Manual
4-27
4
o p e r a t i o n
SETUP MENU LEVEL
3
(CONTINUED)
PVI BAR ON/OFF
BUTTONS
SETTING
Press the
Enter Button
again to move
to the next
menu.
4x
Default
On
Use Up or Down Button to move
between settings
AND
press the Enter Button to accept the
setting and move to the next menu.
OR
press the Display Button to exit without
saving the new setting and to return to
the home display screen.
Off
(No PVI Bar or
parameter/mea-
surement label
displayed)
SMART TONE ON/OFF
The SmartTone feature uses a proprietary algorithm that will provide pulse tones during
excessive motion and low perfusion conditions. The pulse tone is based on an averaged
pulse rate measurement from the proprietary algorithm and may not identify irregular
heart beat patterns when there is excessive artifact present.
The Normal Tone feature uses a proprietary algorithm that will provide pulse tones
during non motion and adequate perfusion conditions. In this mode, the pulse tone
may not sound if excessive artifact is present.
BUTTONS
SETTING
Press the
Enter Button
again to move
to the next
menu.
5x
Default
Off
(Normal Tone)
Use Up or Down Button to move
between settings
AND
press the Enter Button to accept the
setting and move to the next menu.
OR
press the Display Button to exit without
saving the new setting and to return to
the home display screen.
On
(Smart Tone)