Rad-87 Pulse CO-Oximeter Operator’s Manual
1-3
1-3
1
o v e r v i e w
P r o d u c t D e s c r i p t i o n
The Rad-87 Pulse CO-Oximeter Monitor is a noninvasive, arterial oxygen, carboxyhe-
moglobin and methemoglobin saturation, and pulse rate monitor. The Rad-87 features a
multicolored LED display that continuously displays numeric values for SpO
2
, SpCO
®
*, Sp-
Met™*, perfusion index (PI), pleth variability index* (PVI) and pulse rate. It also provides bar
graph displays for quick visual identifi cation of Signal Identifi cation Quality (SIQ
®
), perfusion
index and pleth variability index.
The Rad-87 is available in four models: vertical Rad-87, horizontal Rad-87, vertical Rad-87
with radio and horizontal Rad-87 with radio.
*Optional features: SpCO, SpMet,
PVI
FEATURES
These features are common to Rad-87 monitors:
■
Masimo SET is clinically proven to be the highest sensitivity and specifi city pulse oxime-
ter technology in the world.
■
Rainbow technology continuously and noninvasively measures arterial oxygen satura-
tion (SpO2) and pulse rate (BPM), as well as providing a reliable probe-off detection.
■
Perfusion Index (PI) with trending capability indicates arterial pulse signal strength during
low perfusion.
■
Accurate on cyanotic infants with congenital heart disease when used with an LNOP®
Blue Sensor.
■
Signal
IQ® provides signal identifi cation and quality indication during excessive motion
and low signal to noise situations.
■
FastSat® tracks rapid changes in arterial O2 saturation with high fidelity.
■
Variable pitch provides tonal variance for every 1% change in saturation.
■
Remote alarming interface.
■
Up to 72 hours of trending. (See Section 4,
Trends Setup and Use.
)
■
Allows user to customize the default settings and set the device to retain these settings
through a power off/on cycle.
■
The LCD Display allows the user to view a scrolling marque of (installed) parameter/mea-
surement alarm limits, system information, and wireless radio communication (wireless
radio model only).
OPTIONAL FEATURES
■
Rainbow technology uses 7+ wavelengths of light to continuously and noninvasively mea-
sure carboxyhemoglobin (SpCO), methemoglobin (SpMet), as well as providing a reli-
able probe-off detection.
■
Pleth Variability Index (PVI) may
s
how changes that refl ect physiologic factors such as
vascular tone, circulating blood volume, and intrathoracic pressure excursions. 1
■
Provides a 802.11a/b/g wireless radio which interfaces with compatible systems (wireless
radio model only).
■
Ability to connect to Masimo Patient SafetyNet through a wireless network (wireless radio
model only).
1
The utility of PVI is unknown at this time and requires further clinical studies. Technical factors that may
affect PVI include probe malposition and patient motion.