Rad-87 Pulse CO-Oximeter Operator’s Manual
6-1
6-1
6
t r o u b l e s h o o t i n g
g
T r o u b l e s h o o t i n g
The following chart describes what to do if the Rad-87 system does not operate properly
or fails.
PROBLEM
TYPE
SOLUTION
DEVICE DOES NOT
POWER ON
Low battery/ not plugged
into AC power supply
Connect the AC Power Cord to the
Rad-87 and to an AC outlet. Make sure
that the AC Power Indicator light is on.
BATTERY RUN-TIME
IS SIGNIFICANTLY
REDUCED
Low battery
Contact Technical Services or your local
Masimo representative.
CONTINUOUS
SPEAKER TONE
Internal Failure
Device requires service. Press the
Alarm Silence Button. If alarm continues
to sound, power down device. If the
power button does not turn the device
off, press and hold the Power Button for
5 seconds. Return the device for service.
NO SPEAKER TONE
Pulse tone set to “mute”
Press Up Arrow or Alarm Volume Adjust.
NO ALARM TONE
Alarm Silence Enabled
The System Status Light flashes yellow.
See Section 4,
Alarm Silence
.
BUTTONS FAIL
TO WORK WHEN
PRESSED
Internal Failure
Use auxiliary power down method by
pressing and holding Sensitivity and
Display Buttons simultaneously. Return
for service.
SENSOR OFF
MESSAGE
Sensor not connected to
patient properly.
Sensor is damaged.
Properly reapply the sensor on the
patient and reconnect the sensor to the
unit or patient cable.
If the sensor is damaged, replace the
sensor.
NO SENSOR
MESSAGE
Sensor is disconnected
from patient cable.
Sensor connected upside
down into patient cable.
Check to see if the sensor LED is flash-
ing. Disconnect and reconnect the sen-
sor. If the LED fails to operate, replace
the sensor.