iv
Rad-87 Pulse CO-Oximeter Operator’s Manual
■
Reorient or relocate the receiving device.
■
Increase the separation between the equipment.
■
Connect the equipment into an outlet on a circuit different from that to which the
other device(s) are connected.
■
Consult the manufacturer for help.
■
In order to connect wirelessly to a compatible interface system like Patient
SafetyNet, the Rad-87 should be placed in an environment free from RF shielding,
which could hinder wireless reception.
■
To minimize radio interference, other electrical equipment that emits RF transmis-
sions should not be in close proximity to the RAD-87.
■
Changes or modifications to the wireless radio feature whether intentional or unin-
tentional are prohibited without written approval from Masimo Corporation.
■
The Rad-87 (device with optional radio) transmits real-time sensor connectivity
status indicating a connect and/or disconnect state. If the device is in a failure
mode then the radio power is disabled and an error message is indicated on the
device display. The device does not have a powered state where no information is
transmitted.
■
In accordance with FCC requirements, the Rad-87 (device with optional radio)
must be placed greater then 20 cm from the patient's head.
■
In accordance with FCC requirements, radio accessories on the Rad-87 (device
with optional radio) cannot be attached directly to the patient using any accessory
containing metal components.
■
In accordance with international telecommunication requirements, the frequency
band of 5,150 MHz to 5,250 MHz is only for indoor usage to reduce potential for
harmful interference to co-channel mobile satellite systems.
■
The battery should be adequately charged to ensure backup power in case of AC
power disruption.
■
A functional tester cannot be utilized to assess the accuracy of the Pulse
CO-Oximeter or any sensors.
■
To ensure safety, avoid stacking multiple devices or placing anything on the
device during operation.
■
Ensure the speaker is not covered or the device is placed face-down on bedding
or other sound absorbing surface.
■
To protect against injury from electric shock, follow the directions below:
■
Avoid placing the device on surfaces with visible liquid spills.
■
Do not soak or immerse the device in liquids.
■
Always turn off and disconnect the power cord from the AC power supply
before cleaning the device.
■
Use cleaning solutions sparingly.
SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES (CONTINUED)