10.2.2 Communication Commands:
•
The existing serial communication channel will break if user changes any
of the below serial communication settings through serial communication
command.
•
User has to do re-login to again communicate with COM2 terminal as per
explained procedure.
The communication commands change the parameters of serial communication port COM2 of the Model.
These include SBxx,SPx and SSx.
These commands changes the configuration of COM2 of GPS Master Clock Model MC-1-M while in
configuration mode, hence the serial communication on COM2 terminal between the unit and
HyperTerminal stops.
To reinitialize the serial communication, first disconnect the link and then enter into the settings of the
HyperTerminal and set it same as the present setting of GPS Master Clock Model MC-1-M. Again
connect the link.
•
Command: SBxx [Baud Rate Command]
Applicable Options: SB24 / SB48 / SB96 / SB19 (Please refer ‘H’ Command menu for applicable
meanings of mentioned options)
1.
Suppose that the desired baud rate is 4800, the command should be SB48 (See H: HELP). This
will display following on the HyperTerminal:
-----------------------------------------------------------------------------------------
ENTER COMMAND TO CONFIGURE PARAMETER
OR PRESS ENTER TO CONTINUE.
SB48
-----------------------------------------------------------------------------------------
Now, the communication must have stopped.
2.
Disconnect the link by a right click on Call\disconnect.
3.
Enter the settings by following the path:
File\Properties\Connect To\Configure.