Model: MC-1-M (2U)
Doc. Ref. no. : m05/om/101
Issue no. 14
User’s Manual
Page 16 of 103
Duration of Programmable repetitive event generation output via dry
contact (Per Minute or Hour).
Password Protection
Network Settings:
IP address, Subnet mask, Gateway and SNMP Manager IP addresses
configurable through Telnet remote login.
All network configurations are password protected.
POWER SUPPLY
AC: 90 to 264 V, 47-63 Hz
DC: 125-300 V
Power Consumption: <15W Typical (40W Max)
OPTIONAL
POWER SUPPLY
DC: 18 – 72 V
Power Consumption: 40W Max
PHYSICAL
DIMENSIONS
19” Rack Mountable
Width: 482.6 mm (19”)
Depth: 305 mm (12.0079”)
Height: 2 U – 88 mm (3.46”)
Weight: 4 Kg (approx.)
ENVIRONMENT
Temperature
Operating: 0° C to +55° C
Storage: -20° C to +80° C
Humidity
20 - 90% (Non-condensing)
EXTRA MODULES
(OPTIONAL)
RS232-to-RS485 Converter
LINE AMPLIFIER, SURGE ARRESTOR
Time Distribution Rack (TDR-4)
Time Signal Repeater (TSR-4)
Time Display Unit (TDU-64)
Netser (NGTS to NTP Convertor)
TYPE TEST
Radiated Emission on mains port (RE) – CISPR 22
Electro Static Discharge – IEC 61000-4-2
Radiated Susceptibility – IEC 61000-4-3
Electrical Fast Transient – IEC 61000-4-4
Surge - IEC 61000-4-5
Conducted Susceptibility - IEC 61000-4-6
Power Frequency Magnetic Field – IEC 61000-4-8
High Frequency Disturbance – IEC 61000-4-10
Voltage Dips & Interruption Test – IEC 61000-4-11
Damped Oscillatory Magnetic Field – IEC 61000-4-12
Vibration – IEC 68-2-6
Cold Test – IEC 60068-2-1:2007
Shock Test – IEC 60255-21-2
Dry Heat – IEC 60068-2-2
Damp Heat Steady State Test - IEC 60068-2-30
Di-electric Test, Vibration Test etc.
I
solation
(Withstanding
Between primary terminals* and secondary terminals**:
At least 1500 V AC for 1 minute