15.2 CHANGING CONTACT PULLEYS AND SPRINGS
Note
While mounting, make sure that the center of rotation and the driver
are aligned.
Use genuine spare parts only!
Follow these instructions to change the contact pulley and springs:
Disconnect the peristaltic pump from power supply;
Unscrew nuts of the pump head (wrench size 5.5) ;
1
Pump head nuts
2
Pump head
3
Springs
4
Groove
5
Driver (roll carrier)
6
Collar of the shaft bore
Figure 9 Disassembly of pump head and driver
Remove the pump head from the motor shaft
Now the driver can be removed from the pump head and is ready for maintenance.
The removal of the springs .4 pcs.) away from the driver is easily possible without the aid of
any tools. For this take spring out of the groove near to the shaft bore.
Dismount roller axes and change contact pulleys. Take care that axes are not worn out by the
springs and have damaged the dent at the axes front end. In case of abrasion the axes have to
be changed (see Figure 10).
Figure 10 Check of axes and rolls
The dent prevents
rotation of the axis
new
worn out
Summary of Contents for TEC1 Series
Page 21: ...21 TECHNICAL SPECIFICATIONS ...
Page 22: ...22 DRAWING INDEX TEC1 with Controller Outline Mounting 03P0003 ...
Page 23: ...23 TEC1 available options 03P0005 ...
Page 24: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 47: ...www mc techgroup com SR25 X X 1 01 01 23 Figure 11 Spare parts Drawing No 2435 1 07 0 ...
Page 48: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 49: ...Bellows pump series MP F MP F05 MP F05 R MP F10 MP F10 R Instruction Manual Version 1 04 01 ...