SUPPLY CONNECTIONS (CONT.)
ELECTRICAL CONNECTIONS
When connecting the equipment, please ensure that the supply voltage is identical with the
information provided on the model type plate.
The main AC power supply connection is to a power entry module on the right side of the enclosure.
The TEC1 will be supplied as specified by the customer for either 115 VAC or 230 VAC mains
voltage.
A change in the main voltage will require a change of cooling fan.
Failure to connect a suitable ground potential to the protective conductor terminal may result in a
shock hazard and risk danger of damage.
The power entry module is equipped with a 6A fuse.
Status alarm contacts for indicating any alarm condition should be connected into the equipment control
system. The dry contact output of the status group alarm is located on the terminal block located on the right
outside the enclosure. This contact is rated for the following values: 1A@30VDC, 0.5A@115VAC,
0.3A@60VDC.
Standard single or optional dual water slip sensor inputs will also be located on the terminal block.