24 MP-
4.01 www.mc-techgroup.com
Inspect the following
pump components
Action
Pump
Check pump for external damages and any leakage in regular inter-
vals, at least two times per year.
Capacitor
Check the conditions of the adhesive covers of the vents in regular
intervals. Replace capacitor with damaged adhesive cover.
Bellows and valve plates
Replace at least when performance of the pump decreases.
The valve plates need to be replaced, when they are < 1.6 mm
Connecting rod bearings
Need to be replaced after 20 000 operating hours or 24 month of
operation, whichever occurs first.
Motor bearings
Need to be replaced after 20 000 operating hours or 24 month of
operation, whichever occurs first.
Fittings, connections, inlets Check in suitable intervals.
Replace when damaged with original parts in perfect condition.
Equipotential bonding
Check equipotential bonding between pump enclosure and motor.
The equipotential bonding needs to be lower than 0.3 Ohm. Use lock
washers for the screws.
Additional components, ac-
cessories
During maintenance, any upstream filters, separators or coolers
must be checked for proper functioning.
Note
If there are any damages to the connection rod, e.g. a loose threaded
pin M8 or loose ball bearings, the complete unit with connection rod
and bearings need to be replaced (see chapter 17 for more details)
The eccentric is glued to the motor shaft by using Loctite 270. This
connection is additionally secured by a set screw.
Summary of Contents for TEC1 Series
Page 21: ...21 TECHNICAL SPECIFICATIONS ...
Page 22: ...22 DRAWING INDEX TEC1 with Controller Outline Mounting 03P0003 ...
Page 23: ...23 TEC1 available options 03P0005 ...
Page 24: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 47: ...www mc techgroup com SR25 X X 1 01 01 23 Figure 11 Spare parts Drawing No 2435 1 07 0 ...
Page 48: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 49: ...Bellows pump series MP F MP F05 MP F05 R MP F10 MP F10 R Instruction Manual Version 1 04 01 ...