8 MP-
4.01 www.mc-techgroup.com
Wrong
operation
Warning
Please make sure to install and operate the pump for the in-
tended use described in this instruction manual only.
Warning
The pump can only be operated within the parameters stated in
8 Technical data .
Warning
The pump is not designed for liquids or particles.
It is prohibited to operate the pump with condensate.
The pump head can be electrostatically charged by particles,
liquids and drops of condensate.
Condensate accumulated inside the pump head destroys the
pump.
If it is likely that humidity or condensation occurs, the pump
head needs to be rotated 180° facing downwards. With the
pump head facing downwards, the condensate can leave the
pump head.
Warning
The materials of the pump, which are in contact with the sample
gas, need to be suitable for the used sample gas.
Warning
In order to operate the pump, it needs to be securely installed.
Warning
Install the pump only in protected areas, sheltered from rain
and moisture.
Please make sure to provide sufficient ventilation.
Warning
Do not exeed the permitted maximum pressure within the pump
(gauge pressure).
Do not close the sample gas output. A sufficient sample gas
flow needs to be provided to prevent the pump pressure from
exceeding the permitted maximum pressure value.
A sufficient sample gas flow also prevents the temperature of
the pump head to increase over the permitted temperature
range.
Warning
If components downstream from the pump can block or reduce
the sample gas flow, appropriate measure, e.g. a pressure
valve, needs to be installed to prevent the process pressure
from exceeding the maximum permitted pressure value.
Summary of Contents for TEC1 Series
Page 21: ...21 TECHNICAL SPECIFICATIONS ...
Page 22: ...22 DRAWING INDEX TEC1 with Controller Outline Mounting 03P0003 ...
Page 23: ...23 TEC1 available options 03P0005 ...
Page 24: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 47: ...www mc techgroup com SR25 X X 1 01 01 23 Figure 11 Spare parts Drawing No 2435 1 07 0 ...
Page 48: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 49: ...Bellows pump series MP F MP F05 MP F05 R MP F10 MP F10 R Instruction Manual Version 1 04 01 ...