18 SR25.X-X
|
1.01.01
www.mc-techgroup.com
15.1 CHANGING THE PUMP TUBING
1
Conveying belt
2
S-bolt
3
Tubing set
4
contact pulley
5
springs
Figure 7 Changing the pump tubing
For changing the pump tubing please proceed as follows:
Unplug the pump from the mains voltage. The device needs to be voltage free.
Press conveying belt at the recessed grips and turn S-bolt clockwise up to limit stop;
Take away conveying belt and remove the old tubing set from the guides by pulling on the
tube connectors;
Press the two contact pulleys and check whether the spring pressure is still sufficient, if not,
the contact springs have to be changed (see chapter 15.2);
Put the new tubing set with the tube connectors into the guides of the conveying belt ;
Note
Only the usage of the original tubing set guarantees a proper functionality.
Never lubricate the tube.
Before mounting the pump check all parts for contaminations and clean if
necessary.
Put the conveying belt with the new tubing into the dovetail guide of the pump body;
Press conveying belt at the recessed grips and simultaneously turn the S-bolt anticlockwise
until it snaps;
Switch on pump.
Figure 8 Different pump tube sizes
Summary of Contents for TEC1 Series
Page 21: ...21 TECHNICAL SPECIFICATIONS ...
Page 22: ...22 DRAWING INDEX TEC1 with Controller Outline Mounting 03P0003 ...
Page 23: ...23 TEC1 available options 03P0005 ...
Page 24: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 47: ...www mc techgroup com SR25 X X 1 01 01 23 Figure 11 Spare parts Drawing No 2435 1 07 0 ...
Page 48: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 49: ...Bellows pump series MP F MP F05 MP F05 R MP F10 MP F10 R Instruction Manual Version 1 04 01 ...