Advanced protein features
Chapter 6
Zetasizer Nano accessories guide
Page 6-15
The parameters are described below.
Data
This indicates the data entries that have been entered into the Debye plot table, see
below.
Sample
Sample dn/dc (ml/g)
This is the specific refractive index increment; the change in refractive index as
a function of the change in concentration.
Experiment notes
Used to record specific details about the experiment or calculation performed.
Hydrodynamic radius (nm)
The radius as measured using dynamic light scattering.
Solvent Ref. Index (Refractive index)
The refractive index of the solvent used.
Solvent count rate (kcounts)
The count rate used to calculate the Rayleigh ratio (R
) is the ‘residual’ count
rate, which is derived by subtracting the solvent count rate from the sample
count rate.
Shape Model
The shape model that is used to estimate the radius of gyration from the
hydrodynamic radius, and therefore calculate the angle dependent effects on
KC/RoP for particles of sizes outside the Rayleigh region (Diameter >
~50nm).
System
Wavelength (nm)
The wavelength of the laser used in the Zetasizer Nano instruments, or for the
measurement. Either 632.8nm ‘red’ or 532nm ‘green’ laser wavelengths are
available.
Scattering angle (degrees)
The scattering angle of the optics unit - either 173°, or 90° for the ‘classical’
optics arrangement.
Toluene count rate (kcounts)
The scattering count rate of the toluene reference.
Saving the Debye plot
The parameters and data inputted to produce the plot can be saved by pressing the
Save
button, and then reviewed at a later stage by pressing the
Load
button.
Summary of Contents for Zetasizer Nano Series
Page 2: ......
Page 3: ...Zetasizer Nano accessories guide MAN0487 Issue 1 1 April 2013...
Page 12: ...Chapter 1 Introduction and accessory range Page 1 6 MAN 0487...
Page 42: ...Chapter 2 General cells and cuvettes Page 2 30 MAN 0487...
Page 98: ...Chapter 5 Microrheology Page 5 20 MAN 0487...
Page 125: ......