Chapter 5
Microrheology
Page 5-14
MAN 0487
The
tracer size
measurement is used later in the microrheology measurement
step. The tracer size needs to be measured to ensure the microrheology result
is as accurate as possible.
When finished a dialogue will appear indicating the next step to perform.
Measure tracer zeta potential in dispersant/buffer (or solvent) including
sample
This is termed the
sample
measurement.
Use a zeta potential cell.
Prepare the sample to be used for the measurement, then add small amount of
tracer particles - do this at a concentration similar to that used in the previous
step.
Ensure the tracer and measurement sample are mixed.
A measurement of the zeta potential of the tracer in the measurement sample is
then performed.
Once this measurement is complete the software compares the two measured
zeta potential results and tests to see if
sample
measurement is within a set
percentage of the
tracer
measurement - this is the
acceptable zeta ratio
.
If the ratio is
less than 25%
the measurement process will continue to the next
step otherwise with a dialogue appearing indicating the next step to perform.
If the ratio is
more than 25%
the software will give sample preparation advice
and advise on repeating steps 1 and 2. Refer to the
zeta potential - accepta-
ble zeta
ratio
section.
Summary of Contents for Zetasizer Nano Series
Page 2: ......
Page 3: ...Zetasizer Nano accessories guide MAN0487 Issue 1 1 April 2013...
Page 12: ...Chapter 1 Introduction and accessory range Page 1 6 MAN 0487...
Page 42: ...Chapter 2 General cells and cuvettes Page 2 30 MAN 0487...
Page 98: ...Chapter 5 Microrheology Page 5 20 MAN 0487...
Page 125: ......