57
ELECTROMAGNETIC COMPATIBILIT Y (EMC)
The bed is intended for operation in the electromagnetic environments listed below. The customer or user of the bed must ensure that
it is used in a suitable environment.
Guidelines and manufacturers’ declarations – electromagnetic emissions
Emission measurement
Compliance
Electromagnetic environment directive
RF emissions, CISPR 11
Group 1
The bed uses HF energy exclusively for its internal functions. There-
fore, it produces very low HF emissions and it is unlikely that nearby
electronic devices will be adversely aff ected.
RF emissions, CISPR 11
Class A
The bed is suitable for use in all facilities including residential areas and
those that are directly connected to a public supply network that also
supplies buildings used for residential purposes.
HF emissions IEC 61000-3-2
Class A
Voltage fl uctuations and fl icker IEC 61000-3-3
Compliant
Guidelines and manufacturers’ declarations – electromagnetic resistance
Phenomena
EMC basic
standard or test
procedure
Environments in domestic
health care areas
Compliance level
Electromagnetic environment directives
Static electricity
discharge
IEC 61000-4-2
±8 kV contact
±8 kV contact
The fl ooring material must consist of wood or
concrete. If synthetic materials are used, the
relative humidity must be at least 30 %.
± 2 kV; ± 4 kV; ± 8 kV;
± 15 kV air discharge
± 2 kV; ± 4 kV; ± 8 kV;
± 15 kV air discharge
High-frequency
electromagnet-
ic fi elds
IEC 61000-4-3
10 V/m
10 V/m
Portable and mobile radio devices must not
be used closer to the care bed and its cables
than the recommended distance calculated
using the equation applicable to the frequency
of the transmitter.
80 MHz to 2.7 GHz
80 MHz to 2.7 GHz
80 % AM at 1 kHz