17
The rating plate is located on the underside of the sleeping sur-
face, at the head of the bed on the right. To inspect the rating
plate, raise the head section to the upper position.
1.
Address of manufacturer
2.
Model ID
3.
Serial number
4.
Electrical voltage, frequency, power consumption
5.
Operating time of the motorised adjustment system:
please take note of this information to prevent over-
heating. For example, the bed must not be operated
continuously for more than 2 minutes, and after 2 minutes
of continuous operation, it must be switched off for at
least 18 minutes.
6.
Protection of electrical equipment against splash water
“Only use in dry areas”; protection rating class II (double
insulation, protective insulation)
7.
Indicates the next technical check after delivery in line with
VDE0751-1
8.
Explanation of the symbols used on the rating plate:
Labelling as medical product
Application part type B
Directive 2002 / 96 / EC relating to old electrical
and electronic equipment
Conformity marking in line with the Medical
Device Directive
Protection rating class IPX4 and protection
rating class II
“Only use in dry areas”
“Observe the instruction manual”
9.
Explanation of the symbols used on the rating plate:
Safe working load
Maximum permissible weight of the
resident
Maximum total weight of the
medical product including resident
(Bed weight plus maximum permissible
weight of resident)
10. 2D-Barcode (GS1 Data Matrix) DI + PI = UDI
11. DI (Device Identifi er)
PI (Product Information Number)