46
|
MAINTENANCE
The manufacturer is only liable for the safety and reliability of the
product if it is serviced regularly and used in line with the safety
instructions. If any signifi cant faults are found during maintenance
work which mean the safe operation of the product cannot be
guaranteed, the product must be taken out of use. Maintenance
work must be carried out at least once a year.
Any faults that impair the function and safety of
the care bed must be resolved before the bed is
used again and must be reported to the responsible
person.
Only original replacement parts from Betten Malsch
GmbH are permitted to be used.
Service and maintenance tasks must not be con-
ducted when the bed is occupied. The resident or
the care staff must not conduct maintenance tasks.
Procedure
1. Visual inspection
Check the welded structures for cracked weld seams and for
plastic deformation and wear. The welded structures include the
undercarriage and the sleeping surface with the moving interior
components. Also check that all screw joints are fi rmly connected.
2. Level of protection and functionality check of the side rails
During the functionality check, check whether the side rails can
be easily locked in place and that no impermissible wear or defor-
mation is visible.
The spacing requirements stated in DIN EN 60601-2-52 are used
to check the level of protection. This is shown by the following
diagram and table.
Fig. 1 (dimensions of a single-piece side rail)