The care bed is not suitable for use in HF-shielded rooms
used for magnetic resonance imaging in which high-intensity
EM disturbance variables occur.
The use of this care bed directly next to or in conjunction
with other electrical devices (e. g. stacked) must be avoided
as this can lead to faulty operation. If using the bed in the
manner described above is absolutely necessary, the devices
involved should be subjected to a function test for a longer
period of time, in order to rule out malfunction due to in-
terference.
The use of other accessories, transducers and cables other
than those specifi ed or provided by Betten Malsch GmbH
can cause increased emissions of electromagnetic interfer-
ence or reduce the electromagnetic interference resistance
of the device and lead to faulty operation.
Portable HF communication devices (radio devices) – in-
cluding their accessories, such as antenna cables and ex-
ternal antennas – must not be used closer than 30 cm to
the parts and cables of care beds as designated by Betten
Malsch GmbH. Non-observance can impair the perfor-
mance of the care bed.
The emissions of this device are below the thresholds de-
fi ned by IEC / CISPR 11:2009, Class A and permit the use of
it in industrial environments and hospitals. This device may
not provide adequate protection against wireless services
if used in residential areas (for which class B is normally
required according to CISPR 11). The user may have to
implement remedial measures such as relocating or reposi-
tioning the device.
Servicing and maintenance tasks may not be performed
while the ME device is in use.