Reflex Decay graph
The reflex decay graph shows the reflex decay curve and the half-life time for the selected stimulus.
•
Time line in seconds
•
Stimulus on/stimulus off markers
•
Deflection curve for the selected stimulus
•
Half-life marker
Reflex Decay results
Use the buttons in the Reflex section on the front panel to navigate between the measurements. Your selection is high-
lighted in the Reflex results table.
The results table shows the results related to the currently selected curve:
•
Stimulus side
•
Stimulus frequency
•
Half-life time
If the system cannot automatically analyze a reflex measurement, a
Not Tested
indication is shown.
On-screen selections
Adjust Marker
Press this soft-button so that you can adjust the half-life time marker by using the
dial.
Hide Marker
/
Show Marker
If, for example, the result is ambiguous, you can hide a half-life time marker.
Alternatively, you can show the marker.
Done
When you have adjusted the marker, press the
Done
soft-button.
The results table is updated with the selected value.
9.8
Manual Tympanometry
In Manual Tympanometry you can control the pump pressure manually.
The default setup available in the tympanometry test is designed for any of the tympanometry tests you can
perform. Adjust the settings to suit your purposes.
Preparing for testing
1. Do as described in
.
2. Ask the patient to sit very still and quiet during the test, without moving head or jaw.
3. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
The probe lights up in the ear color to indicate that it is activated.
4. Press the
Special
button and select
Manual Tymp
.
5. The diagnostic probe is selected automatically when you select the
Manual Tymp
test.
66
Otometrics - MADSEN Zodiac
9 Testing with Zodiac Diagnostic or Clinical - Stand-Alone
Summary of Contents for zodiac
Page 1: ...MADSEN Zodiac Reference Manual Doc No 7 50 1610 EN 02 Part No 7 50 16100 EN...
Page 21: ...4 2 Bottom view A Probe connections B Label Otometrics MADSEN Zodiac 21 4 Views of Zodiac...
Page 156: ...jB jB App 2 5 2 G App 2 5 3 B G B G Y B G B G 156 Otometrics MADSEN Zodiac App 2 MADSEN Zodiac...
Page 157: ...App 2 5 4 226 Hz 226 Hz 200 daPa Otometrics MADSEN Zodiac 157 App 2 MADSEN Zodiac...
Page 206: ...152 151 151 152 150 152 206 Otometrics MADSEN Zodiac Index...