3. Check to see if the sound channels of the probe
tip are blocked. You can fit a new probe tip on
the probe, or use the supplied probe tip
flossing thread to clean the sound channels.
Caution
•
Never clean the sound channels in
the probe body, as this may cause damage
to the probe.
A.
Sound channels
4. Fit the probe tip on the probe, and twist the
probe ring clockwise to lock the probe tip in
place on the probe.
Probe check
If you have fitted a new probe tip, do a probe check. See
17.4.5
Fitting a new probe tip
If the probe tip has been contaminated or blocked by cerumen or vernix, fit a new probe tip on the probe.
Procedure
A.
Probe body
B.
Probe ring
C.
Probe tip
122
Otometrics - MADSEN Zodiac
17 Service and maintenance