D. Tympanometry results
The results table shows the results related to the currently selected curve:
•
Probe
(the probe tone in Hz)
•
TPP
(Tympanometric Peak Pressure)
•
ECV
(Equivalent Ear Canal Volume)
•
SA
(Static Peak Admittance), or
SC
(Static Peak Compliance) when volume equivalent units are used
•
TW/Ratio
(Tympanometric Width/Tympanometric Ratio). Describes the steepness of the curve.
•
Type
(Jerger types A, As, Ad, B, C, D and E denote the shape of the 226 Hz curve). You can set the type to be
determined automatically, and you can subsequently change it manually, if necessary.
E. On-screen selections
Curve selection
Press the
Previous Curve
or
Next Curve
softbutton to select the desired curve in
the Reflex Screening and Tympanometry result tables.
Manually adjusting the TPP
Press the
Edit TPP
softbutton, use the dial to move the marker to the desired
TPP, and press the
Done
softbutton.
Changing tympanogram curve type
Press the
Edit TPP
softbutton. Press the
Change Type
softbutton until the
desired type is shown, and press the
Done
softbutton.
Deleting a curve
Select the curve you wish to delete, and press the
Delete
button.
To replace a single measurement, delete a curve and adjust the measurement
selections to redo the single measurement.
Otometrics - MADSEN Zodiac
47
8 Testing with Zodiac Quick Check - Stand-Alone
Summary of Contents for zodiac
Page 1: ...MADSEN Zodiac Reference Manual Doc No 7 50 1610 EN 02 Part No 7 50 16100 EN...
Page 21: ...4 2 Bottom view A Probe connections B Label Otometrics MADSEN Zodiac 21 4 Views of Zodiac...
Page 156: ...jB jB App 2 5 2 G App 2 5 3 B G B G Y B G B G 156 Otometrics MADSEN Zodiac App 2 MADSEN Zodiac...
Page 157: ...App 2 5 4 226 Hz 226 Hz 200 daPa Otometrics MADSEN Zodiac 157 App 2 MADSEN Zodiac...
Page 206: ...152 151 151 152 150 152 206 Otometrics MADSEN Zodiac Index...