Use the dial or the arrow buttons to select another test protocol, if needed.
•
Press the soft-button below the next column to select this column, and scroll to any setting you wish to view or
change.
How to change test settings
1. Start by selecting a protocol.
2. Select the center column.
3. Scroll to the setting you wish to view or change.
4. Select the right-hand column to change the setting.
5. Scroll to the value you wish to use and press the
Select
soft-button.
6. Repeat steps 2 to 5 to change other settings, if needed.
Using test settings temporarily
•
If you only wish to use the changed protocol settings temporarily, select the left-hand column, scroll to
Apply Tem-
porarily
and press the
OK
soft-button. This will exit the menu system.
Saving test settings in the protocol
Before you can save test settings in the protocol, you must unlock the protocol: Select
Device Settings..
>
Protocol Set-
tings
, and set to
Allowed
. If you switch off Zodiac after changing and saving the test settings, the protocol will be locked
again.
•
If you wish to save the changed protocol settings permanently, select the left-hand column, scroll to
Save to [Protocol
name]
, and press the
Save
soft-button.
Canceling changed test settings
•
If you wish to cancel the settings you have changed, select the protocol and press the
Reload
soft-button.
6.4
Special tests
The Diagnostic and Clinical versions of Zodiac can perform a number of special immittance tests, such as ETF-P and Manual
Tympanometry.
Accessing special tests
When you press the
Special
button, the available tests are shown.
You can access the
Probe Check
function from the
Special Tests
menu.
Loading a special test
1. Press the
Special
button.
2. Scroll to select the special test you wish to use.
3. Press the
OK
soft-button.
Exiting a special test
Press the
Special
button to exit the special test. Zodiac returns to Tympanometry mode.
34
Otometrics - MADSEN Zodiac
6 MADSEN Zodiac test controls
Summary of Contents for zodiac
Page 1: ...MADSEN Zodiac Reference Manual Doc No 7 50 1610 EN 02 Part No 7 50 16100 EN...
Page 21: ...4 2 Bottom view A Probe connections B Label Otometrics MADSEN Zodiac 21 4 Views of Zodiac...
Page 156: ...jB jB App 2 5 2 G App 2 5 3 B G B G Y B G B G 156 Otometrics MADSEN Zodiac App 2 MADSEN Zodiac...
Page 157: ...App 2 5 4 226 Hz 226 Hz 200 daPa Otometrics MADSEN Zodiac 157 App 2 MADSEN Zodiac...
Page 206: ...152 151 151 152 150 152 206 Otometrics MADSEN Zodiac Index...