H-6
Specifications
DPX-IQ Operator’s Manual
Mains frequency
50/60 Hertz
IEC 601-2-7 6.1j)3
Maximum apparent resistance of supply mains
0.8 Ohms
IEC 601-2-7 6.1j)4
Required over-current releases
15 Amp dedicated service
IEC 601-2-7 6.1j)5
Heat dissipative components
X-ray tube dissipates 305W max. into
surrounding air through forced air
convection. Flow rate: 36 m
3
/h (approx.)
Temp. rise of air stream 25º C (approx.)
IEC 601-2-7 6.1t)
Allowable high voltage supplies
Spellman X2112/X2113/ rev. K and higher.
Bertran 2411P and 2411N rev. A and
higher. LUNAR p/n 0311 and 0312.
IEC 601-2-7 6.8.1
and 50.2.101-102
Allowable tube head assemblies
LUNAR model 2288 or equivalent
IEC 601-2-7 6.8.1
and 50.2.101-102
Original language of accompanying documents
English
IEC 601-2-7 6.8.1
Maximum contiuous kV, mA at nominal rated kV
76 kV, 4 mA
IEC 601-2-7 6.8.2 1)
Maximum intermittent kV, mA at maximum kV
76 kV, 5 mA
IEC 601-2-7 6.8.2 1)
Maximum continuous kV, mA at maximum mA
61 kV, 5 mA
IEC 601-2-7 6.8.2 2)
Maximum intermittent kV, mA at maximum mA
76 kV, 5 mA
IEC 601-2-7 6.8.2 2)
Continuous kV, mA for maximum electric output
power
76 kV, 4 mA
IEC 601-2-7 6.8.2 3)
Intermittent kV, mA for maximum electric output
power
76 kV, 5 mA
IEC 601-2-7 6.8.2 3)
Nominal electric power
0.4 kW
IEC 601-2-7 6.8.2 4)
Reference current time product
7.89 mAs. Parameters: 76 kV, 2.63 mA,
3 seconds.
IEC 601-2-7 6.8.2 5)
Nominal shortest irradiation times
3 seconds.
IEC 601-2-7 6.8.2 8)
Repetition rate for loading during tests
No specific wait period was imposed.
Time between tests was approximately
20 seconds.
IEC 601-2-7
50.104.4
Method of X-ray tube voltage measurement
Voltage divider in high voltage power
supply.
IEC 601-2-7
50.106.1
Method of X-ray tube current measurement
Shunt resistor in high voltage supply
return line.
IEC 601-2-7
50.106.2
X-ray tube assembly reference axis
Line normal to the tube port, centered on
tube port as shown in figure H-1.
IEC 601-2-7 6.1j)5)
50.107.1
Reference loading conditions
1.09 x 10
6
Joules, 4 mA, 76 kV for 1 hour.
IEC 601-1-3
Leakage radiation was measured at the following
loading factors.
750mA, 76 kV for 1 hour.
IEC 601-1-3
Focal spot to Image Receptor distance
58.1 cm
IEC 601-1-3
Attenuation equivalence of patient support table.
0.7 mm Al
IEC 601-1-3
Table H-1: X-ray generator technical information. (continued)
Summary of Contents for DPX-IQ
Page 6: ...vi DPX IQ Operator s Manual This page left blank intentionally...
Page 12: ...xii Table of Contents DPX IQ Operator s Manual...
Page 54: ...4 8 Default Settings DPX IQ Operator s Manual Figure 4 3 Analysis Results report 4 6d...
Page 56: ...4 10 Default Settings DPX IQ Operator s Manual This page left blank intentionally...
Page 68: ...A 12 AP Spine DPX IQ Operator s Manual This page left blank intentionally...
Page 79: ...DPX IQ Operator s Manual Femur B 11 Figure B 12 Position Neck ROI...
Page 80: ...B 12 Femur DPX IQ Operator s Manual This page left blank intentionally...
Page 100: ...D 10 Total Body DPX IQ Operator s Manual This page left blank intentionally...
Page 130: ...G 14 Orthopedic Hip DPX IQ Operator s Manual Figure G 15 Transverse Histogram screen...
Page 160: ...J 6 Precision and Accuracy DPX IQ Operator s Manual This page left blank intentionally...
Page 166: ...This page left blank intentionally...