DPX-IQ Operator’s Manual
Forearm
E-1
Appendix E Contents
E.1 Forearm Image Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
E.2 Acquisition Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3
E.3 Forearm Image Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-5
Revised 1/99C
Appendix E:Forearm Image Acquisition and Analysis
E.1
Forearm Image Acquisition
The procedure that follows gives the basic steps necessary to complete a forearm
image acquisition. Refer to section E.2 for more information.
1.
Refer to section 2.3.3 to add or select a patient.
2.
At the Forearm Acquisition screen, select
#
. The Verify Values window is shown.
3.
Record or examine the following information:
•
Forearm Length
—the length of the forearm to be scanned.
•
Patient Side
—the arm to be scanned, left or right. (LUNAR recommends the
patient’s non-dominant arm.)
•
Scan Mode
—150
µ
A is the only scan mode available. Select
!
when the Scan
Mode window is shown.
•
Scan Width
—width of the arm to be scanned. Make sure the width is large
enough to scan the patient's forearm.
•
Scan Length
—record Total, Distal, Shaft, or Manual. (LUNAR recommends
Total scan length.)
4.
Select
/
. The Forearm Acquisition screen is shown.
5.
Select
!
. The scan arm moves to the HOME position.
6.
Remove the scan pad and set the forearm positioner on the scan table.
E
Forearm
Summary of Contents for DPX-IQ
Page 6: ...vi DPX IQ Operator s Manual This page left blank intentionally...
Page 12: ...xii Table of Contents DPX IQ Operator s Manual...
Page 54: ...4 8 Default Settings DPX IQ Operator s Manual Figure 4 3 Analysis Results report 4 6d...
Page 56: ...4 10 Default Settings DPX IQ Operator s Manual This page left blank intentionally...
Page 68: ...A 12 AP Spine DPX IQ Operator s Manual This page left blank intentionally...
Page 79: ...DPX IQ Operator s Manual Femur B 11 Figure B 12 Position Neck ROI...
Page 80: ...B 12 Femur DPX IQ Operator s Manual This page left blank intentionally...
Page 100: ...D 10 Total Body DPX IQ Operator s Manual This page left blank intentionally...
Page 130: ...G 14 Orthopedic Hip DPX IQ Operator s Manual Figure G 15 Transverse Histogram screen...
Page 160: ...J 6 Precision and Accuracy DPX IQ Operator s Manual This page left blank intentionally...
Page 166: ...This page left blank intentionally...