DPX-IQ Operator’s Manual
Orthopedic Hip
G-1
Appendix G Contents
G.1 Orthopedic Hip Image Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-1
G.2 Orthopedic Hip Image Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-4
G.3 Analysis Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-6
Revised 1/99C
Appendix G:Orthopedic Hip Image Acquisition and Analysis
G.1
Orthopedic Hip Image Acquisition
1.
Refer to section 2.3.3 to add or select a patient.
2.
At the Orthopedic Hip Acquisition screen, select
#
to examine values. Record or
examine the information that follows:
•
Porous Length
—Record the length of the porous coating. If you do not know
the porous length use the default value of 0.
•
Implant Length
—Record the length of the patient’s implant. If you do not
know the implant length use the default value of 100.
•
Patient Side
—Record the hip being scanned, left or right.
•
Mode
—Highlight the applicable mode for patient thickness (table G-1).
Select
!
.
•
Scan Width
—Use the default value of 150 if you do not have a specified
scan width.
•
Scan Length
—Use the default value of 100 if you do not have a specified
scan length.
G
Orthopedic Hip
Summary of Contents for DPX-IQ
Page 6: ...vi DPX IQ Operator s Manual This page left blank intentionally...
Page 12: ...xii Table of Contents DPX IQ Operator s Manual...
Page 54: ...4 8 Default Settings DPX IQ Operator s Manual Figure 4 3 Analysis Results report 4 6d...
Page 56: ...4 10 Default Settings DPX IQ Operator s Manual This page left blank intentionally...
Page 68: ...A 12 AP Spine DPX IQ Operator s Manual This page left blank intentionally...
Page 79: ...DPX IQ Operator s Manual Femur B 11 Figure B 12 Position Neck ROI...
Page 80: ...B 12 Femur DPX IQ Operator s Manual This page left blank intentionally...
Page 100: ...D 10 Total Body DPX IQ Operator s Manual This page left blank intentionally...
Page 130: ...G 14 Orthopedic Hip DPX IQ Operator s Manual Figure G 15 Transverse Histogram screen...
Page 160: ...J 6 Precision and Accuracy DPX IQ Operator s Manual This page left blank intentionally...
Page 166: ...This page left blank intentionally...