G-8
Orthopedic Hip
DPX-IQ Operator’s Manual
Figure G-8. Edge marker placement left femur
G.3.2
Examine Reference Point
Make sure the Reference Point (
➊
) is immediately next to the lesser trochanter
(figure G-9). Make sure the horizontal line of the Reference Point is in the center of
the lesser trochanter and is parallel to the femoral shaft. In addition, the vertical line of
the Reference Point touches the curve of the lesser trochanter.
NOTE: Only Standard Analysis uses a Reference Point.
Figure G-9. Reference Point