Lunar DPX-IQ Operator'S Manual Download Page 55

DPX-IQ Operator’s Manual

Default Settings

4-9

4.3

Select Language

The Select Language option lets you set the language shown on the screen and on 
results printouts. Seven languages are given: English, French, German, Japanese, 
Spanish, Italian, and Portuguese. English is the default language.

NOTE: These options are available only if you installed the language 

options.

1.

At the Options screen, select 

*

. The Select Language menu is shown.

2.

Highlight the necessary language, and select 

!11

. The program shows the language 

on the Options screen.

4.4

Select Reference Population

The LUNAR program lets you compare patient bone mineral density (BMD) to a 
database of reference information. You can choose from several national databases. 
Use of the reference population comparisons is fully at the discretion of the 
clinician.

NOTE: These options are available only if you installed the reference 

population options.

1.

At the Options screen, select 

+

. The Select Reference Group menu appears.

2.

Highlight the necessary regional database and select 

!

. The program compares 

BMD of the AP spine, lateral spine, femur, total body, and forearm to information 
from this population until you select a different one.

Summary of Contents for DPX-IQ

Page 1: ...rope N V avenue R Vandendriessche 18 9 1150 Brussels Belgium 32 2 7770910 32 2 7791039 Fax LUNAR GmbH Horbeller Strasse 33 D 50858 K ln Germany 49 2234 954590 49 2234 273153 Fax LUNAR France 1 avenue...

Page 2: ...isassemble any Program or any part thereof or merge any part of any Program into another computer program 2 Rent sell sublicense assign or otherwise transfer the Programs to a third party In the event...

Page 3: ...ity 1 thoroughly familiarize yourself with the patient s history and 2 install the scanner in an environment effectively shielded from other sources of external radiation Federal Law restricts this de...

Page 4: ...mergency stop button to routinely abort a scan Using the DPX IQ Operator s Manual Use this manual to help you complete daily acquisition and analysis procedures Chapter 1 System Overview gives scanner...

Page 5: ...X DPX alpha DPX L We hereby declare that the above described product conforms to the requirements of the European Council Directive 93 42 EEC Medical Devices Directive Annex II This declaration is bas...

Page 6: ...vi DPX IQ Operator s Manual This page left blank intentionally...

Page 7: ...1 10 1 2 9 Scattered Radiation 1 10 1 3 System Maintenance 1 13 1 3 1 Clean the Scan Table Environment 1 13 1 3 2 Archive Scan Files 1 13 1 3 3 Test the Emergency Stop Button 1 13 1 4 Preventive Maint...

Page 8: ...Archive Scan Files 2 11 2 6 1 Procedure to Archive Patient Scan Files 2 11 Change Archive Destination and Directory 2 12 Chapter 3 Quality Assurance 3 1 Quality Assurance Tests 3 1 3 1 1 Complete a QA...

Page 9: ...Space Markers A 6 Examine Labels A 6 Examine Intervertebral Space Markers A 6 A 3 2 Examine Baseline and Edge Markers A 7 The Line Profile A 7 Examine Baselines A 8 Examine Edge Markers A 9 A 3 3 Com...

Page 10: ...easure the Forearm Length E 3 E 2 2 Record the Scan Length E 4 E 2 3 Position the Forearm Positioner E 4 Forearm Position E 5 E 3 Forearm Image Analysis E 5 E 4 Analysis Information E 8 E 4 1 Examine...

Page 11: ...3 H 3 Functional Specifications H 3 H 3 1 General Specifications H 3 H 3 2 Maximum Scan Area Long x Transverse H 3 H 3 3 Programs H 4 H 4 Environmental Specifications H 4 H 5 Power Specifications H 5...

Page 12: ...xii Table of Contents DPX IQ Operator s Manual...

Page 13: ...Revised 1 99C Chapter 1 System Overview 1 1 Scan Table Assembly NOTE DO NOT attempt servicing the scan table assembly Please call the LUNAR Customer Support Department or your LUNAR distributor Figure...

Page 14: ...er monitor The position switches status lights and emergency stop button are on the control panel figure 1 2 Figure 1 2 Scan arm control panel Position switches The position switches move the scan arm...

Page 15: ...the laser It is not likely that you can receive more than 25 of the maximum permissible x ray dose from the DPX IQ scanner However some facilities choose to use personnel monitors Refer to your city...

Page 16: ...between the scan arm and the scan table At the end of a Lateral Spine acquisition a pinch point can be made between the patient s legs and scan arm The scan arm makes a pinch point on both ends of the...

Page 17: ...yearly scattered radiation dose for the operator is approximately 166 67 Gy per year This assumes an estimated 10 patients per day 2 scans per patient and average time of 2 minutes per scan and 50 wee...

Page 18: ...3 00 12 x 12 270 810 77 1 2 x 1 2 Total Body Screening 76 0 15 54 x 150 290 44 0 1 4 8 x 19 2 Total Body Fast 76 0 15 54 x 150 580 87 0 2 4 8 x 9 6 Total Body Medium 76 0 15 54 x 150 1160 174 0 3 4 8...

Page 19: ...rays are off and not being generated at the same time the X ray On light on the scan arm control panel is off When this symbol appears with a green background the scanner is ready to begin making X ra...

Page 20: ...ssembly This label gives collimator assembly information It is located on the collimator and foot panel of the scanner p n 5531 Rev D For USA Complies with Radiation Performance Standards in 21 USCFR...

Page 21: ...Emergency stop button again to reset the system 3 Select The system turns off the x ray and the scan arm moves to the HOME position The Options screen is shown NOTE If there is a hardware problem DO...

Page 22: ...computer equipment and accessories connected to the scanner must comply with IEC 950 Safety of information technology equipment including electrical business equipment and must conform with IEC 801 5...

Page 23: ...PX IQ scanner was run at 76 kV and 4 75 mA EMERGENCY POWER ON X RAY SHUTTER LASER Back Front Left Right DPX IQ R Detector 1 Gy hr Tube Head Foot end View Water 1 Gy hr Gy hr Gy hr Gy hr Gy hr Gy hr Gy...

Page 24: ...canner was run at 76 kV and 0 75 mA DPX IQ scanner was run at 76 kV and 4 75 mA EMERGENCY POWER ON X RAY SHUTTER LASER Back Front Left Right DPX IQ R 91 4 cm 30 cm 50 8 cm 25 4 cm 30 5 cm 50 8 cm 30 c...

Page 25: ...makes free space on your hard drive Refer to section 2 6 for more information 1 3 3 Test the Emergency Stop Button Test the Emergency Stop button once a month Refer to the procedure that follows 1 Sta...

Page 26: ...ir the components and scan table Doing so voids all current warranty and service contracts 1 4 2 Daily Quality Assurance Procedure Complete system Quality Assurance procedures daily Make sure each QA...

Page 27: ...ion in this manual the term select is the act of pressing a key to operate the program 2 1 1 Options Screen When you start the LUNAR program an options screen is shown figure 2 1 The options that foll...

Page 28: ...lect Reference Population to access the reference databases necessary to compare BMD results Exit to leave the LUNAR program and go back to the Windows desktop Figure 2 1 Options screen 2 1 2 Common K...

Page 29: ...e x ray beam such as clothing with zippers snaps buckles and buttons Require patients to wear a jogging suit to the exam or give them an examination gown when they arrive Radionuclides and Radiopaque...

Page 30: ...tient position Step 4 Locate Scan Start Position You must use the laser light to locate the scan start position for the detector A correct scan start position makes sure the program accurately scans t...

Page 31: ...lbs or 136 4 kg Select 0 after each record NOTE To record patient height use whole numbers and make sure decimals greater than 0 5 are rounded to the next highest number 5 Select or to record patient...

Page 32: ...isition screen is shown Change the scan values and select to start the scan 4 Continue the acquisition procedure End the Scan To end a scan select then select two times During a femur scan make sure 2...

Page 33: ...o change the image Type The options for Type are Bone and Soft Bone shows a skeletal image and Soft shows a soft tissue image Soft is an option for the Total Body scan type only Filtering Filtering sm...

Page 34: ...g may require you to complete manual adjustments to the scan image such as ROI or edgemarker positions Refer to the applicable appendix for the scan type A thru G for procedures to analyze and if nece...

Page 35: ...by the reference standard deviation SD A Z score of 2 means the patient is 2 SDs below the mean age matched value of the reference population Print a DICOM Report optional If you have purchased the D...

Page 36: ...t Print select to print the Chronological Results screen for one patient 2 5 2 Graph and Print Patient Results 1 At the Chronological Results screen select the necessary scan type For example to graph...

Page 37: ...an IQNET system 1 At the Options screen select The Database Utilities screen is shown 2 Select The Patient Directory screen shows scan files related to patient name 3 Select one of the options that f...

Page 38: ...een you can change the drive destination and directory used to archive patient files 1 At the Database Utilities screen select 2 Move the cursor to the line that follows Archive Drive a ar_data 3 Type...

Page 39: ...If a test fails two times call the LUNAR Customer Support Department or your LUNAR distributor DO NOT try to scan a patient until a successful QA test is complete 3 1 1 Complete a QA Test Use the stan...

Page 40: ...o the correct results that follow The detector peak and the high keV and low keV peaks are no more than 50 units above or below the previous peak settings Each Functional Measurement has a spillover o...

Page 41: ...lect to begin a QA procedure QUALITY ASSURANCE RESULTS X Ray Voltage kVp 76 0 Date 01 01 98 X Ray Current uA 150 0 System 0000 AMPLIFIER COUNTS COUNTS SETTING LOW keV HIGH keV 265 59862 1882 315 66818...

Page 42: ...pical density range and size of a normal human spine The spine phantom figure 3 3 consists of vertebrae L1 through L4 Parts of vertebrae T12 and L5 are landmarks used by the DPX IQ program during anal...

Page 43: ...er to the QA printout before this scan if you think the values are not correct Call LUNAR Customer Support or your Distributor 8 Select to go back to the AP Spine Acquisition screen 9 To scan the enca...

Page 44: ...mate start position and the program turns on the laser light 14 Position the laser light figure 3 6 in the middle of L5 and in the center of the phantom Select to start the scan L5 L4 L3 L2 L1 T12 C A...

Page 45: ...A for more information 1 From the AP Spine Options screen select The Patient Directory screen is shown Select the Spine Phantom Select 2 Select the phantom scan file for analysis Select The Mandatory...

Page 46: ...ecision Over Time The Chronological Results screen lets you show graph and print results of spine phantom scans over time in chronological order 1 At the AP Spine Options screen select The Database Ut...

Page 47: ...ify the date and time you want to print scan analysis or Quality Assurance results 4 1 1 Print Manager Options At the Options screen select to access the Print Manager screen The Print Manager screen...

Page 48: ...which relates to the day of the week necessary to print the reports Select 0 NOTE If you select Daily the program prints reports each day at the selected time 5 Select then to go back to the Options s...

Page 49: ...information which let you personalize analysis printouts a typical heading can show the name of your institution department or physician s name and address of your institution NOTE The Report Heading...

Page 50: ...the type of report that is printed after scan analysis NOTE Option 4 sBMD is given only for AP Spine L2 L4 Total Femur and Femur Neck analysis Display Country Field Display Country Field lets you sho...

Page 51: ...y Options 1 At the Report Heading screen select The Printout Display screen is shown 2 Highlight the value to be changed then select to turn the option ON or to turn the option OFF Each option is give...

Page 52: ...e categorized as osteoporosis The footnote on the Results Report explains the T scores associated with each category The T score represents the patient s BMD result above or below a reference Young Ad...

Page 53: ...because of smaller body sizes the values for U S women of Asian descent are similar to Caucasian values Age Regression Line Age Regression Line shows the line in the center of the Age Matched Regressi...

Page 54: ...4 8 Default Settings DPX IQ Operator s Manual Figure 4 3 Analysis Results report 4 6d...

Page 55: ...m shows the language on the Options screen 4 4 Select Reference Population The LUNAR program lets you compare patient bone mineral density BMD to a database of reference information You can choose fro...

Page 56: ...4 10 Default Settings DPX IQ Operator s Manual This page left blank intentionally...

Page 57: ...and select to go back to the AP Spine Acquisition screen 4 Select The scan arm moves to the HOME position 5 Help the patient lie on the scan table Select The scan arm and detector move to the approxi...

Page 58: ...comfortable Figure A 1 Patient position 7 Position the laser light figure A 2 approximately 5 cm below the patient s navel and in the same longitudinal plane as the patient s midline Figure A 2 Laser...

Page 59: ...to the scan image such as ROI or edgemarker positions The procedure that follows gives the basic steps necessary to complete an AP spine image analysis Refer to section A 3 for more information 1 Refe...

Page 60: ...Results screen 6 Select The Analysis Results screen is shown Refer to figure A 4 which shows the two possible graphic styles you may see For AP spine analysis you can choose to display a WHO Fracture...

Page 61: ...DPX IQ Operator s Manual AP Spine A 5 Figure A 5 AP Spine Analysis Report Figure A 6 AP Spine Ancillary Report...

Page 62: ...e program labeled the necessary vertebrae To change labels complete the procedure that follows 1 At the Spaces Labels screen select The Select Labels window is shown 2 Use the arrow keys to select the...

Page 63: ...ure A 8 Spine image and profile The Line Profile The Line Profile shows the information that follows figure A 9 Baseline The baseline is the horizontal line at the bottom of the profile The points bel...

Page 64: ...correctly change the baseline ONLY if you are sure it is not correct Figure A 10 illustrates correct baseline position When you change the baseline the program automatically changes the edge markers...

Page 65: ...nwanted soft tissue and is positioned on the baseline NOTE Calcification fat deposits and dense substances can cause unwanted soft tissue points to be shown on the Line Profile Figure A 11 Incorrect E...

Page 66: ...t analysis Complete one of the steps that follow Select to let the software position the ROIs Select to not change the ROIs To remove or add ROIs select option or NOTE If you select the Bone Results s...

Page 67: ...tor s Manual AP Spine A 11 7 Select The Analysis Results screen is shown 8 Select to save changes 9 Select and record the number of Analysis Results and or Ancillary reports necessary Select to print...

Page 68: ...A 12 AP Spine DPX IQ Operator s Manual This page left blank intentionally...

Page 69: ...2 for a left femur or 3 for a DualFemur scan 4 Select The Mode window is shown Refer to table B 1 to make sure the scan mode you select is correct for scan region thickness Highlight the applicable m...

Page 70: ...h below the horizontal line on the scan pad Figure B 1 Patient position 8 The DPX IQ unit includes one of the two foot braces shown in figures B 2 and B 3 Follow the instructions for your model s brac...

Page 71: ...le Rotate the ENTIRE leg until the foot is positioned at an angle against the foot brace 10 Select The scan arm and detector move to the approximate starting position and the program turns on the lase...

Page 72: ...eater trochanter A message is shown which asks you if you want to continue the scan on the same femur If it appears that less than 25 lines of the scan are acquired above the greater trochanter select...

Page 73: ...t tissue of leg Femur If necessary use two rice bags to get 18 75 cm of tissue equivalent material for a 3000 A scan The top rice bag is tight against the leg to prevent air gaps Let the patient hold...

Page 74: ...rogram is acquiring the left or right femur Refer to section 2 3 5 for more information Abort between left and right femur scans You can select to abort the scan while the scan arm and detector are mo...

Page 75: ...Trochanteric ROIs Refer to section B 4 2 for a description of these ROIs 5 Select The Bone Results screen is shown 6 Use the arrow keys to highlight the region to be shown on the Analysis Results scr...

Page 76: ...B 8 Femur DPX IQ Operator s Manual Figure B 8 Femur Analysis Report Figure B 9 Femur Ancillary Report...

Page 77: ...e scan was acquired incorrectly because of a bad starting location aborting too soon etc If the Bone Results Screen shows one of these messages refer to the procedure in section B 1 to complete the sc...

Page 78: ...Neck ROI position Position the Neck ROI Complete the procedure that follows only if the program did not position the Neck ROI correctly 1 Select from the Bone Results screen 2 Select to move to expan...

Page 79: ...DPX IQ Operator s Manual Femur B 11 Figure B 12 Position Neck ROI...

Page 80: ...B 12 Femur DPX IQ Operator s Manual This page left blank intentionally...

Page 81: ...in the scan region Patients more than 30 cm thick can not be scanned by the detector 3 Examine the other scan values and select 1 to go back to the Lateral Spine Acquisition screen 4 Select The scan...

Page 82: ...e positioner and below the knees Make sure the patient s back is flat against the positioner and adjust the strap so the patient does not roll forward during the scan Position a foam wedge between the...

Page 83: ...c crest If the detector position is not correct select to stop the scan Refer to section 2 3 5 Figure C 3 Correct lateral spine scan image 11 Continue the scan past the top of L2 Select to stop the sc...

Page 84: ...d of the keyhole slots Figure C 4 Lateral positioner C 3 Lateral Spine Image Analysis NOTE In most cases the software will accurately analyze the scan image However unusual anatomy or incorrect patien...

Page 85: ...ey are wrong 4 Select The Auto Analysis screen is shown Make sure the ROI positions are correct 5 Select The Bone Results screen is shown 6 Use the arrow keys to highlight the region to be shown on th...

Page 86: ...C 6 Lateral Spine DPX IQ Operator s Manual Figure C 6 Lateral Spine Analysis Report Figure C 7 Lateral Spine Ancillary Report...

Page 87: ...gram the Body ROI for L2 is labeled B2 L3 is labeled B3 and L4 is labeled B4 Middle ROI M The Middle ROI contains the middle part of the Body ROI The program supplies results for the middle part of th...

Page 88: ...e they are between the vertebral bodies and located at the lowest point of bone density as shown on the histogram If a space marker is not correct use the arrow keys to select the marker and select Re...

Page 89: ...s At the Auto Analysis screen select to verify profiles The Profiles screen shows the edge markers on the scan image figure C 11 Make sure that each profile is positioned along the edges of the verteb...

Page 90: ...n of each ROI figure C 12 The ROI contains only bone The top and bottom edges of the ROI are in the center of the vertebral endplates The endplates are the light grey regions at the top and bottom of...

Page 91: ...ion Refer to section D 2 for more information NOTE Make sure the patient removes all attenuation materials before you start the acquisition procedure 1 Refer to section 2 3 3 to add or select a patien...

Page 92: ...ith palms against the scan table Figure D 1 Patient position 6 Select to start the scan The scan arm and detector move to the start position 7 Make sure the patient s head and feet are shown in the sc...

Page 93: ...ith step 5 3 Select Use the options shown on the screen to change cut positions 4 Select The Bone Results Screen is shown 5 Use the arrow keys to highlight the region to be shown on the Analysis Resul...

Page 94: ...the bottom of the Analysis Results screen call LUNAR Customer Support 7 Select to save changes 8 Select and record the necessary number of Analysis Results and or Ancillary reports figures D 4 D 5 an...

Page 95: ...DPX IQ Operator s Manual Total Body D 5 Figure D 5 Total Body Analysis Report page 2 Figure D 6 Total Body Ancillary Report 3 6y...

Page 96: ...show tissue quantitation results in the Body Composition section of the Ancillary results printout The tissue information combines fat and lean soft tissue Other tissue data in the report combines sof...

Page 97: ...ow the chin Ribs There is a left and right rib cut shown on the scan image Position each cut as close to the spine as possible Do not include the spine in the rib region Arms There is a left and right...

Page 98: ...esults for patients who weigh less than 75 kg Extended Research Analysis is recommended for patients who weigh more than 75 kg It is necessary to carefully position cuts for Extended Research Analysis...

Page 99: ...processes information from the current cut positions and the Bone Results screen is shown If specific anatomical sites select and refer to section D 3 1 to position the cuts When finished select The p...

Page 100: ...D 10 Total Body DPX IQ Operator s Manual This page left blank intentionally...

Page 101: ...Verify Values window is shown 3 Record or examine the following information Forearm Length the length of the forearm to be scanned Patient Side the arm to be scanned left or right LUNAR recommends the...

Page 102: ...1 Forearm position 9 Select The scan arm and detector move to the approximate start position and the program turns on the laser light 10 Position the laser light figure E 2 in the center of the patie...

Page 103: ...r the scan stops select to continue the scan for 15 20 more lines Select 13 After the scan remove the patient s arm from the positioner Select to go back to the Forearm Options screen Figure E 3 Corre...

Page 104: ...ngth Shaft scans the 33 shaft site Distal scans the distal end of the forearm and includes the Ultradistal site the 5 mm site and the 10 site Manual lets you record the necessary scan length Figure E...

Page 105: ...ses the software will accurately analyze the scan image However unusual anatomy or incorrect patient positioning may require you to complete manual adjustments to the scan image such as ROI or edgemar...

Page 106: ...s forearm and select If no continue with step 6 6 Make sure the Reference Point and ROI positions are correct If changes are necessary use the options shown on the screen to position the reference po...

Page 107: ...DPX IQ Operator s Manual Forearm E 7 Figure E 9 Forearm Analysis Report Figure E 10 Forearm Ancillary Report...

Page 108: ...e Region Labels To change the region labels select The Select Labels window is shown and gives the label combinations Each line contains a combination of the Region labels in the order shown on the sc...

Page 109: ...baseline down Please refer to section A 3 2 for more information on baselines Examine Edge Markers Make sure the edge markers are positioned above the baseline to include only high density points Refe...

Page 110: ...he radius and ulna meet as you move up the forearm to the wrist Figure E 13 Reference point and ROI position E 4 4 Examine ROI Heights The height of each ROI is shown on the right side of the Auto Ana...

Page 111: ...positioned on the scan image where the ulna and radius are 5 mm apart figure E 15 this distance is given on the right side of the Auto Analysis screen 10 ROI The recommended height of the 10 ROI is 1...

Page 112: ...E 12 Forearm DPX IQ Operator s Manual Total ROI The Total ROI contains the 33 ROI in figure E 16 the Ultradistal ROI and the area between them Figure E 16 Total ROI...

Page 113: ...ide record the hand to be scanned left or right Mode The Mode window shows the 150 A detail mode Select to set the mode Scan Width select Manual for scan width Measure the width of the hand at its wid...

Page 114: ...ight figure F 2 in the center of the wrist immediately distal to the ulna styloid process Figure F 2 Laser light position 9 Select to start the scan Monitor the first few lines to make sure the detect...

Page 115: ...sis 1 Refer to section 2 4 1 to select a scan file for analysis 2 At the Bone Results screen select The Auto Analysis screen is shown 3 Select The Analysis Results screen is shown figure F 4 4 Select...

Page 116: ...omments 1 See appendix on precision and accuracy Statistically 68 of repeat scans will fall within 1 SD 0 02 g cm2 BMD 1 7 Young Adult 2 Age Matched 3 Region g cm2 T Z TOTAL 0 422 Reference Data Not A...

Page 117: ...ecord or examine the information that follows Porous Length Record the length of the porous coating If you do not know the porous length use the default value of 0 Implant Length Record the length of...

Page 118: ...G 1 Knee positioner 8 Secure the foot of the leg to be scanned to the straight vertical side of the foot brace Make sure the patient s toes point up figure G 2 Refer to the information that follows t...

Page 119: ...inue the scan until at least 5 lines show above the top of the greater trochanter NOTE If white lines show on the edges of the scan image there is not enough soft tissue around the scan region There m...

Page 120: ...r positions The procedure that follows gives the basic steps necessary to complete an orthopedic hip image analysis Refer to section G 3 for more information 1 Refer to section 2 4 1 to select a patie...

Page 121: ...Orthopedic Hip G 5 7 Select and record the necessary number of Analysis Results and or Ancillary reports figures G 5 and G 6 Select to print the reports Figure G 4 Bone Results screen Figure G 5 Ortho...

Page 122: ...lect from the Auto Analysis screen to examine orthopedic hip edge markers the Profiles screen shows the scan image and Line Profile figure G 7 The Line Profile shows areas of the analysis region the p...

Page 123: ...Figure G 7 Line Profile Examine Edge Markers Refer to the information that follows to make sure the edge marker positions are correct figure G 8 Pelvis Edge If the pelvis is shown on the scan image th...

Page 124: ...diately next to the lesser trochanter figure G 9 Make sure the horizontal line of the Reference Point is in the center of the lesser trochanter and is parallel to the femoral shaft In addition the ver...

Page 125: ...is ROIs figure G 10 ROI C is positioned by the program so its bottom edge is at the top of the lesser trochanter ROI D is immediately below ROI C The bottom edge of ROI C and the top edge of ROI D tou...

Page 126: ...settings ROIs 2 and 6 are immediately above ROIs 3 and 5 There is no space between the bottom of ROI 2 and the top of ROI 3 In addition there is no space between the bottom of ROI 6 and the top of RO...

Page 127: ...ning ROIs contain bone not affected by the porous coating After the boundaries are set the program divides the implant length equally between ROIs 2 and 6 and ROIs 3 and 5 The program positions ROI 4...

Page 128: ...ample if you select an Offset value of 4 mm the program shows a set of edge markers 4 mm from the implant 1 At the Profiles screen select The number following Offset below the line profile is highligh...

Page 129: ...tudinal Histogram figure G 14 shows the density of bone next to the implant Use this histogram to monitor bone remodeling along the entire length of the implant The program calculates BMD values for t...

Page 130: ...G 14 Orthopedic Hip DPX IQ Operator s Manual Figure G 15 Transverse Histogram screen...

Page 131: ...Specifications H 4 H 5 Power Specifications H 5 H 6 X ray Generator Specifications H 5 H 7 Compatible Components H 10 Revised 1 99C Appendix H Specifications H 1 Space Requirements Room dimensions mu...

Page 132: ...mponent specifications H 2 1 Printer Your system may include one of the printers that follow For your printer s specifications refer to product literature HP LaserJet 6L 17 5 H x 8 2 W x 15 3 D 44 3 c...

Page 133: ...eneral Specifications Focal spot to image receptor distance is 58 4 cm Attenuation equivalence of patient support table is 0 7 mm AI H 3 2 Maximum Scan Area Long x Transverse NOTE The maximum scan len...

Page 134: ...a around the scan table Make sure there is a minimum clearance of 12 30 5 cm at the head and foot ends of the scan table at least 6 15 2 cm for the arm side and 18 45 7 cm for the operator side Ambien...

Page 135: ...label on your scanner to determine which power specifications apply to your scanner Bertan Spellman 100 Range 90 110 VAC 50 60 Hz 10A P N 0311 and 0312 115 Range 103 127 VAC 50 60 Hz 10A 230 240 Range...

Page 136: ...2 2 Continuous kV mA for maximum electric output power 76 kV 4 mA IEC 601 2 7 6 8 2 3 Intermittent kV mA for maximum electric output power 76 kV 5 mA IEC 601 2 7 6 8 2 3 Nominal electric power 0 4 kW...

Page 137: ...al anode input power 361 Watts for an equivalent Anode input power of 361 W IEC 613 1989 Maximum anode heat content 6000 Joules IEC 613 1989 Anode heating and cooling curves Refer to figure H 2 IEC 61...

Page 138: ...9 Nominal X ray tube voltage 76 kV Anode to Cathode 38 kV Anode to Earth 38 kV Cathode to Earth IEC 613 1989 Single load rating 361 W 4 75 mA 76 kV for up to 4 min 59 sec IEC 613 1989 Serial load rati...

Page 139: ...H 6 4 Laser Specifications Table H 4 gives the specifications for the LUNAR laser Table H 4 Laser specifications Output Power 1mW Wavelength 670nm Frequency Drift 0 25nm C Laser Structure Gain Guided...

Page 140: ...e contact LUNAR for more information For customers located internationally make sure the computer is certified to local requirements such as IEC 950 The computer must meet the requirements that follow...

Page 141: ...3 2GB 32MB 233MHz w Windows 95 5579 Controller DPX IQ ACER America2 Computer Models 2021 2GB 16MB 166MHz 4761 2181 6 4GB 32MB 233MHz 5524 2341 8MB RAM 16 MB RAM 1 6GB 32MB 133MHz black 4115 4013 4491...

Page 142: ...TV40P200X2112 0311 0312 Bertan6 Models 2411AP 2411AN 0312 0311 Tube Head Assembly LUNAR X Ray Tube Head Assembly 2288 1 Dell Computer Corporation Austin TX 2 ACER America Corporation San Jose CA 3 CIT...

Page 143: ...s In the past reference values obtained with a different instrument LUNAR DP3 have been shown to be virtually identical in different white populations US Australia UK Spain Brazil Argentina 1 6 Variat...

Page 144: ...I 3 list the mean BMD values used as DPX s AP spine reference data BMD values are presented in grams per centimeter squared g cm2 in ten year intervals The regions of interest are AP spine L2 L4 and l...

Page 145: ...entheses 32 36 49 Age n Neck Ward s Trochanter Total 0 12 0 13 0 11 0 12 20 29 739 0 998 0 945 0 796 1 007 30 39 1067 0 973 0 888 0 782 0 998 40 49 1746 0 946 0 839 0 782 0 979 50 59 1799 0 881 0 753...

Page 146: ...uivalent BMD52 Table I 6 lists the mean BMD used in LUNAR s total body reference data for females Table I 7 lists the mean BMD used in LUNAR s total body reference data for males BMD values are given...

Page 147: ...7 Mean male BMD values for total body sites SD in g cm2 for each site is listed below site name in parenthesis 32 Age n Total Body Arms Legs Trunk Pelvis Spine 0 08 0 08 0 09 0 07 0 10 0 14 20 29 35...

Page 148: ...0 984 1 345 0 986 1 204 1 146 40 49 83 1 210 0 995 1 339 0 990 1 196 1 166 50 59 86 1 232 1 000 1 351 1 001 1 196 1 179 60 69 236 1 203 0 977 1 332 0 996 1 164 1 203 70 79 151 1 177 0 937 1 306 0 974...

Page 149: ...0 139 45 65 665 Ap Spine 1 71 0 0113 X Age 0 32 0 174 65 166 Ap Spine 1 16 0 0026 X Age 0 20 0 170 20 45 171 Lateral Spine 0 895 0 0047 X Age 0 25 0 118 45 65 125 Lateral Spine 1 190 0 0104 X Age 0 46...

Page 150: ...25 and 100 kg I 4 1 Young Adult The Young Adult value compares a patient s BMD with the expected peak bone mass for a subject of the same sex and nationality The Young Adult age ranges for each scan s...

Page 151: ...eacock M Slemenda C Johnston CC Weaver CM 1995 Peak bone mass in young women J Bone Miner Res 10 711 715 8 Bell NH Gordon L Stevens J Shary JR 1995 Demonstration that bone mineral density of the lumba...

Page 152: ...determinants of peak bone mass and subsequent rates of bone loss In Osteoporosis I Christiansen C Arnaud CD Nordin BEC Parfitt AM Peck WA Riggs BL eds Proceedings of the Copenhagen International Symp...

Page 153: ...spinal bone mineral density measured laterally a normal range for UK women Br J Radiol 67 157 161 36 Truscott JG Simpson D Fordham JN Compilation of national bone densitometry reference data in Ring E...

Page 154: ...in women a longitudinal study over 15 years Calcif Tissue Int 49 90 94 48 Kr ger H Laitinen K 1992 Bone mineral density measured by dual energy x ray absorptiometry in normal men Eur J Clin Nutr 22 45...

Page 155: ...g a spine or femur The DPX has shown a better precision than Gd153 based DPA about 01 g cm2 error on spine and femur phantoms Refer to tables J 1 and J 2 Measurements with the DPX system were repeated...

Page 156: ...0 015 Table J 2 Short term 5 days precision of DPX for regions of total body scans Intrascanner n 34 and interscanner n 37 results for single skeletons in vitro and results for young subjects in vivo...

Page 157: ...o intercept Accuracy for vertebral area was further validated by comparing the measured areas of three vertebral bodies L2 L4 in nine encapsulated spine phantoms using the DPX Figure J 1 Actual vs mea...

Page 158: ...DP4 scanner in 11 young subjects The correlation between results was high r 0 96 SEE 0 02 g cm2 The regression line was congruent with the identity line figure J 3 Figure J 3 Comparison of DPA and DE...

Page 159: ...valent to 25 cm patient thickness Corrections for scatter and hardening reduce this small thickness dependence at higher thicknesses figure J 5 Figure J 5 Effect of tissue thickness on results J 3 Ref...

Page 160: ...J 6 Precision and Accuracy DPX IQ Operator s Manual This page left blank intentionally...

Page 161: ...ay line profile 4 4 Femur B 6 Forearm E 5 Hand F 3 Lateral Spine C 4 Orthopedic Hip G 4 Results 2 8 Scan image 2 8 Total Body D 3 AP Spine Acquisition A 1 Analysis A 3 Baselines A 8 Body ROI A 10 Edge...

Page 162: ...itor scan B 5 Neck ROI position B 9 Rice bags B 5 Scan modes B 1 Filtering 2 7 Forearm 10 ROI E 11 33 ROI E 10 5mm ROI E 11 Acquisition E 1 Analysis E 5 Baseline E 9 Bone labels E 8 Edge markers E 9 F...

Page 163: ...ormation Database mini menus 4 4 Display country field 4 4 Options screen 2 1 Orthopedic Hip Acquisition G 1 Analysis G 4 Edge markers G 7 Extened Gruen analysis G 11 Gruen analysis G 11 Gruen Analysi...

Page 164: ...1 4 Operator 1 3 Patient 1 4 Pregnancy restrictions 2 3 Radiation 1 5 Symbols 1 6 System 1 3 Scaling factor 2 7 Scan image Adjust 2 7 Analyze 2 8 Scan modes 1 5 Scan table 1 1 Components 1 1 Control p...

Page 165: ...DPX IQ Operator s Manual Index 5 X X ray generator H 5 X ray tube Lunar 2288 H 7 H 8 Lunar 8022 H 7 Y Young Adult Results defined 2 9 Young adult BMD comparison I 8 Display on printout 4 5...

Page 166: ...This page left blank intentionally...

Reviews: