B-10
Femur
DPX-IQ Operator’s Manual
➌
the ROI, ideally, should not contain part of the ischium. Part of the ischium can
be contained in the ROI if there is a clear separation between the ischium and the
femoral neck.
➍
the ROI does not include any portion of the trochanter.
Figure B-11. Correct Neck ROI position
Position the Neck ROI
Complete the procedure that follows
only
if the program did not position the Neck
ROI correctly:
1.
Select
$
from the Bone Results screen.
2.
Select
#
to move,
$
to expand, or
%
to rotate the ROI to position the ROI
(figure B-12): make sure the ROI is
➊
perpendicular to the axis of the femoral neck, and
➋
near the proximal (top) end of the femoral neck.
3.
Select
+
then
!
. The program finds the area that contains the lowest bone density
and positions the other three ROIs.
➊
➊
➋
➌
➍
Summary of Contents for DPX-IQ
Page 6: ...vi DPX IQ Operator s Manual This page left blank intentionally...
Page 12: ...xii Table of Contents DPX IQ Operator s Manual...
Page 54: ...4 8 Default Settings DPX IQ Operator s Manual Figure 4 3 Analysis Results report 4 6d...
Page 56: ...4 10 Default Settings DPX IQ Operator s Manual This page left blank intentionally...
Page 68: ...A 12 AP Spine DPX IQ Operator s Manual This page left blank intentionally...
Page 79: ...DPX IQ Operator s Manual Femur B 11 Figure B 12 Position Neck ROI...
Page 80: ...B 12 Femur DPX IQ Operator s Manual This page left blank intentionally...
Page 100: ...D 10 Total Body DPX IQ Operator s Manual This page left blank intentionally...
Page 130: ...G 14 Orthopedic Hip DPX IQ Operator s Manual Figure G 15 Transverse Histogram screen...
Page 160: ...J 6 Precision and Accuracy DPX IQ Operator s Manual This page left blank intentionally...
Page 166: ...This page left blank intentionally...