Chapter 3:Quality Assurance
3.1
Quality Assurance Tests
Complete a Quality Assurance (QA) test each morning before you scan a patient. The
procedure lasts 10–15 minutes and makes sure quality scans are acquired. Save all
QA printouts.
If a QA test fails, the message that follows is shown:
Unsuccessful Quality Assurance
Repeat procedure suggested
Complete the QA tests again. If a test fails two times, call the LUNAR Customer
Support Department or your LUNAR distributor. DO NOT try to scan a patient until a
successful QA test is complete.
3.1.1
Complete a QA Test
Use the standard to complete a QA test (the standard is a block of tissue-equivalent
material with three bone-simulating chambers of known bone mineral content).
1.
At the Options screen, select
%
. The System Quality Assurance screen is shown.
2.
Remove the pad from the scan table.
3.
Position the calibration standard on the table (figure 3-1):
➊
Put the standard inside the rectangle on the scan table.
➋
Position the standard so the brass-colored part of the standard faces upward and
is toward the front of the scan table. DO NOT move the standard after the
test starts.
3
Quality Assurance
Summary of Contents for DPX-IQ
Page 6: ...vi DPX IQ Operator s Manual This page left blank intentionally...
Page 12: ...xii Table of Contents DPX IQ Operator s Manual...
Page 54: ...4 8 Default Settings DPX IQ Operator s Manual Figure 4 3 Analysis Results report 4 6d...
Page 56: ...4 10 Default Settings DPX IQ Operator s Manual This page left blank intentionally...
Page 68: ...A 12 AP Spine DPX IQ Operator s Manual This page left blank intentionally...
Page 79: ...DPX IQ Operator s Manual Femur B 11 Figure B 12 Position Neck ROI...
Page 80: ...B 12 Femur DPX IQ Operator s Manual This page left blank intentionally...
Page 100: ...D 10 Total Body DPX IQ Operator s Manual This page left blank intentionally...
Page 130: ...G 14 Orthopedic Hip DPX IQ Operator s Manual Figure G 15 Transverse Histogram screen...
Page 160: ...J 6 Precision and Accuracy DPX IQ Operator s Manual This page left blank intentionally...
Page 166: ...This page left blank intentionally...