DPX-IQ Operator’s Manual
Specifications
H-7
Figure H-1. Reference axis and target angles for
tube head assembly
H.6.2
LUNAR 8022 X-ray Tube
Table H-2 gives LUNAR 8022 X-ray tube information.
Figure H-2. Anode heating/cooling curves
Table H-2: LUNAR 8022 X-ray tube technical information.
Nominal anode input power
361 Watts for an equivalent
Anode input power of 361 W
IEC 613/1989
Maximum anode heat content
6000 Joules
IEC 613/1989
Anode heating and cooling curves
Refer to figure H-2.
IEC 613/1989
Anode target material
Tungsten
IEC 601-2-28
Reference axis
Refer to figure H-3.
IEC 601-2-28
Target angle
78° (reference to normal)
IEC 601-2-28
Nominal focal spot values
0.5
IEC 336/1982
Maximum useful voltage
95 kVp
Not Applicable
Maximum filament current
2.2 Amperes
Not Applicable
Summary of Contents for DPX-IQ
Page 6: ...vi DPX IQ Operator s Manual This page left blank intentionally...
Page 12: ...xii Table of Contents DPX IQ Operator s Manual...
Page 54: ...4 8 Default Settings DPX IQ Operator s Manual Figure 4 3 Analysis Results report 4 6d...
Page 56: ...4 10 Default Settings DPX IQ Operator s Manual This page left blank intentionally...
Page 68: ...A 12 AP Spine DPX IQ Operator s Manual This page left blank intentionally...
Page 79: ...DPX IQ Operator s Manual Femur B 11 Figure B 12 Position Neck ROI...
Page 80: ...B 12 Femur DPX IQ Operator s Manual This page left blank intentionally...
Page 100: ...D 10 Total Body DPX IQ Operator s Manual This page left blank intentionally...
Page 130: ...G 14 Orthopedic Hip DPX IQ Operator s Manual Figure G 15 Transverse Histogram screen...
Page 160: ...J 6 Precision and Accuracy DPX IQ Operator s Manual This page left blank intentionally...
Page 166: ...This page left blank intentionally...