Teledyne Lumenera Lt Series
Connecting the Camera
•
59
4.2.2
LED Status Indicator
The Lt Series has one multicolor LED located on the back panel, to provide a simple visible
indication of camera state, as described below.
Lt-Ux0x-xxxxx, Lt-Ux1x-xxxxx
Lt-Ux2x-xxxxx
Description
LED is off
LED is off
No power to the camera
Flashing red
Initialization sequence
Steady orange
Device driver failed to load
Flashing orange few times per
second
Steady red
Firmware failed to load
Flashing orange once per 2
seconds
Steady red
FPGA failed to load
Steady green
Steady blue
Camera ready / operational
Dimmer green
Flashing green
Camera is streaming or fast frames mode is
enabled
Dimmer green (snapshot)
Flashing green
LED will be dimmer green for duration of
exposure/LED will flash on frame acquisition
Flashing orange/green
(case 1) Camera is connected to USB 2.0
ports and does not have external power
connected
(case 2) USB 3 port is not supplying enough
camera power
Steady red and flashing blue
Fatal Error. If the camera does not reboot
itself contact Technical Support
4.3
Preventing Operational Faults due to ESD
Camera installations which do not protect against ESD (electrostatic
discharge) may exhibit operational faults. Problems such as random
camera resets and random loss of connections may all be solved by
proper ESD management.
The camera, when used with USB 3 power or a simple power supply, is
not properly connected to earth ground and therefore is susceptible to
ESD caused problems.
Teledyne Lumenera has performed ESD testing on cameras using an 8 kilovolt ESD generator
without any indication of operational faults. The two following methods, either individually or
together will prevent ESD problems.
•
Method 1: Use a shielded/grounded power supply that connects to the ground pin of the I/O
connector. The camera body is now properly connected to earth ground and can withstand ESD
of 8 kilovolts.
•
Method 2: Mount the camera on a metallic platform with a good connection to earth ground.
Summary of Contents for Lt-C1610
Page 80: ...74 Feature Reference Teledyne Lumenera Lt Series...
Page 98: ...92 Technical Specifications Teledyne Lumenera Lt Series 6 2 2 Models Lt Ux1x xxxxx...
Page 99: ...Teledyne Lumenera Lt Series Technical Specifications 93 6 2 3 Models Lt Ux20 xxxxx...
Page 108: ...102 Technical Specifications Teledyne Lumenera Lt Series Lt Ux2x xxxxx models...
Page 111: ...Teledyne Lumenera Lt Series Technical Specifications 105 Lt Ux2x xxxxx models...
Page 112: ...106 Technical Specifications Teledyne Lumenera Lt Series Lt Ux2x xxxxx models...
Page 113: ...Teledyne Lumenera Lt Series Technical Specifications 107 Lt Ux2x xxxxx models...
Page 117: ...Teledyne Lumenera Lt Series Technical Specifications 111 Certificate 2 of 3...
Page 118: ...112 Technical Specifications Teledyne Lumenera Lt Series Certificate 3 of 3...
Page 140: ...134 Troubleshooting Teledyne Lumenera Lt Series 8 7 2 Chart 2...
Page 141: ...Teledyne Lumenera Lt Series Troubleshooting 135 8 7 3 Chart 3...
Page 142: ...136 Troubleshooting Teledyne Lumenera Lt Series 8 7 4 Chart 4...