Gocator 2300 & 2880 Series
Protocols • Gocator Protocol • 317
Restore
The Restore command uploads a backup file to the connected sensor and then restores all sensor files
from the backup.
The sensor must be reset or power-cycled before the restore operation can be completed.
Field
Type
Offset
Description
length
32u
0
Command size including this field, in bytes.
id
16u
4
Command identifier (0x1014).
length
32u
6
Data length.
data[length]
byte
10
Data content.
Command
Field
Type
Offset
Description
length
32u
0
Reply size including this field, in bytes.
id
16u
4
Reply identifier (0x1014).
status
32s
6
Reply status. For a list of status codes, see
Reply
Restore Factory
The Restore Factory command restores the connected sensor to factory default settings.
The command erases the non-volatile memory of the main device.
This command has no effect on connected Buddy sensors.
Note that the sensor must be reset or power-cycled before the factory restore operation can be
completed.
Field
Type
Offset
Description
length
32u
0
Command size including this field, in bytes.
id
16u
4
Command identifier (0x4301).
resetIp
8u
6
Specifies whether IP address should be restored to default:
0 – Do not reset IP
1 – Reset IP
Command
Field
Type
Offset
Description
length
32u
0
Reply size including this field, in bytes.
id
16u
4
Reply identifier (0x4301).
status
32s
6
Reply status. For a list of status codes, see
Reply
Summary of Contents for Gocator 2300 Series
Page 1: ...USER MANUAL Gocator 2300 2880 Series Firmware version 4 3 x xx Document revision D ...
Page 20: ...Gocator 2300 2880 Series Getting Started System Overview 20 ...
Page 184: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 184 Measurement Panel ...
Page 189: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 189 ...
Page 199: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 199 Measurement Panel ...
Page 208: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 208 else Output_Set Volume 0 ...
Page 389: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 389 Dimensions Envelope ...
Page 392: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 392 Dimensions Envelope ...
Page 394: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 394 Envelope ...
Page 396: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 396 Dimensions ...
Page 397: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 397 Envelope ...
Page 399: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 399 Envelope ...
Page 401: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 401 Dimensions ...
Page 402: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 402 Envelope ...
Page 405: ...Gocator 2300 2880 Series Specifications Gocator 2880 Sensor 405 Dimensions ...
Page 406: ...Gocator 2300 2880 Series Specifications Gocator 2880 Sensor 406 Envelope ...