Gocator 2300 & 2880 Series
Protocols • Gocator Protocol • 315
Field
Type
Offset
Description
length
32u
0
Reply size including this field, in bytes.
id
16u
4
Reply identifier (0x4518).
status
32s
6
Reply status. For a list of status codes, see
Reply
Schedule Analog Output
The Schedule Analog Output command schedules an analog output event. The analog output must be
configured to accept software-scheduled commands and be in the Running state. For information on
setting up the analog output, see
Field
Type
Offset
Description
length
32u
0
Command size including this field, in bytes.
id
16u
4
Command identifier (0x4519).
index
16u
6
Index of the output. Must be 0.
target
64s
8
Specifies the time (clock ticks) or position (encoder ticks) of when
the event should happen.
The target value is ignored if
is set to false.
(
Scheduled
is unchecked in
Analog
in the
Output
panel.) The
output will be triggered immediately.
value
32s
16
Output current (micro-amperes).
Command
Field
Type
Offset
Description
length
32u
0
Reply size including this field, in bytes.
id
16u
4
Reply identifier (0x4519).
status
32s
6
Reply status. For a list of status codes, see
Reply
The analog output takes about 75 us to reach 90% of the target value for a maximum change,
then roughly another 40 us to settle completely..
Ping
The Ping command can be used to test the control connection. This command has no effect on sensors.
Field
Type
Offset
Description
length
32u
0
Command size including this field, in bytes.
id
16u
4
Command identifier (0x100E).
timeout
64u
6
Timeout value (microseconds).
Command
Summary of Contents for Gocator 2300 Series
Page 1: ...USER MANUAL Gocator 2300 2880 Series Firmware version 4 3 x xx Document revision D ...
Page 20: ...Gocator 2300 2880 Series Getting Started System Overview 20 ...
Page 184: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 184 Measurement Panel ...
Page 189: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 189 ...
Page 199: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 199 Measurement Panel ...
Page 208: ...Gocator 2300 2880 Series Gocator Web Interface Measurement 208 else Output_Set Volume 0 ...
Page 389: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 389 Dimensions Envelope ...
Page 392: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 392 Dimensions Envelope ...
Page 394: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 394 Envelope ...
Page 396: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 396 Dimensions ...
Page 397: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 397 Envelope ...
Page 399: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 399 Envelope ...
Page 401: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 401 Dimensions ...
Page 402: ...Gocator 2300 2880 Series Specifications Gocator 2300 Series 402 Envelope ...
Page 405: ...Gocator 2300 2880 Series Specifications Gocator 2880 Sensor 405 Dimensions ...
Page 406: ...Gocator 2300 2880 Series Specifications Gocator 2880 Sensor 406 Envelope ...