Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 14.
CHANGED SETTINGS PROGRAMMING
WARNING Screen
Models 100, 101, 102
For Models 103, 104, and 105 Generators, the device parameters are
programmed and verified as a group during a programming event;
therefore, these Pulse Generators are not susceptible to partial
programming. If an interruption occurs during a programming step for
these Pulse Generators, the Programming Software displays a warning
message indicating that the procedure failed and allows the user to retry or
cancel the programming operation (see Figure 15 on page 26). If the user
selects
Retry
, the Programming Software will re-program the Pulse
Generator. If the user selects
Cancel
, the Programming Software will
return to the PARAMETER screen where the user can re-program the
Pulse Generator to desired settings (see Figure 3 on page 15).