42 of 92 — 26-0008-1100/5
(Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
Standby" will be displayed at the top of the screen for the Models 100,
101, and 102 Pulse Generators.
Figure 25.
MAGNET MODE DIAGNOSTICS Screen
To obtain accurate information on Models 100, 101, and 102 Pulse
Generators from the device diagnostics, program the Pulse Generator to
values shown in Table 11 for both normal and magnet modes.
Table 11.
Minimum Normal Mode and Magnet Mode
Diagnostic Settings (Models 100, 101, and 102)
4.2.2.6.
Generator Diagnostics
Note
The optional Generator Diagnostics (Pulse Generator Test), as shown in
Figure 26 on page 43, is performed when the System Diagnostics fails
during the surgical implant procedure.
Models 100, 101, 102
Models 103, 104, 105
Output Current / Magnet Current
0.75 mA
Frequency
15 Hz
Signal / Magnet ON Time
30 sec