Physician’s Manual — VNS Therapy Programming Software (8.1)
c.
Tighten the setscrew(s) until the hex screwdriver clicks.
d. Perform the Generator Diagnostics.
If the Generator Diagnostics indicates “HIGH” Lead
impedance, contact LivaNova at 1 (866) 882-8804 (U.S. and
Canada) or +1 (281) 228-7330 (Worldwide).
If the Generator Diagnostics indicates “OK” Lead impedance,
perform these steps:
1) Back out the setscrew(s) and remove the test resistor.
2) Insert the Lead connector pin(s) into the Lead
receptacle(s) in the Pulse Generator.
3) Tighten the setscrew(s) until the hex screwdriver clicks.
4) If the System Diagnostics continues to indicate “HIGH”
Lead impedance, contact LivaNova at 1 (866) 882-8804
(U.S. and Canada) or +1 (281) 228-7330 (Worldwide).