26-0008-1100/5
(Worldwide)
— 19 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 9.
PULSE DISABLED and BATTERY STATUS
INDICATOR - EOS WARNING Screen
Models 103, 104, 105
Table 1 shows the various indications of Pulse Generator battery level and
the associated recommendations.
Table 1.
Pulse Generator Battery Level Indications and
Recommendations
Pulse
Generator
Model #
Estimated %
Battery Power
Remaining
Battery
Icon
Displayed
Pulse Generator Battery Status /
Recommendations
100/101/102
Battery
measurement not
supported by
Generator
None
displayed
Near EOS = No
The Pulse Generator battery level is good.
No warning message is displayed from the
Programming Software and no special
attention is required.
100/101/102
Battery
measurement not
supported by
Generator
None
displayed
Near EOS = Yes
The Pulse Generator is nearing End of
Service (EOS). The Programming Software
will display a warning message indicating
this status upon completion of interrogation
or Diagnostics testing. It is recommended
that the Pulse Generator be replaced as
soon as possible.