Physician’s Manual — VNS Therapy Programming Software (8.1)
6.
Output Status:
For the Models 100, 101, and 102 Pulse Generators,
this parameter indicates whether the program current is being
delivered at the specified value. Table 5 shows the possible values and
interpretation for this parameter. This parameter is not displayed for
the Models 103, 104, and 105 Pulse Generators.
Table 5.
Device Diagnostic Screen Status Readings for
Output Status
7.
Output Current:
For the Models 100, 101, and 102, this parameter
indicates the stimulation output current applied during the diagnostic
test. Table 6 shows the possible values and interpretation for this
parameter. For the Models 103, 104, and 105 this parameter indicates
whether the programmed current is being delivered during the
diagnostic tests.
8.
Current Delivered:
On the Models 103, 104, and 105 Pulse
Generators, this parameter indicates the output current delivered
during the execution of the diagnostic test. This parameter is not
displayed for Models 100, 101, 102 Pulse Generators.
Output Status
Status Reading
Significance
*****
1. MAGNET MODE diagnostics found Pulse
Generator in STANDBY MODE. Use the
magnet to reactivate stimulation. Retry
diagnostics.
2. NORMAL MODE diagnostics found Pulse
Generator in STANDBY MODE. Retry
diagnostics.
OK
Current is being output at the programmed
value.
LIMIT
Programmed current is possibly not being
delivered at the specified level (possibly
limited by battery voltage, lead impedance,
or other reason). Reprogram the Pulse
Generator to a lower output current and a
wider pulse width.