Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 45.
High Lead Impedance, Low Lead Impedance, or
Low Output Current on an Interrogation or
Diagnostic Test at Follow-up Visit for Models 103,
104, and 105 Pulse Generators
Error message encountered:
Lead Impedance “LOW”
”A LOW Lead Impedance has
been detected...”
Perform a Systems Diagnostics.
Record results.
System Diagnostics Results
Lead Impedance “HIGH”
Output Current “OK”
Possible Causes:
Lead discontinuity
Lead disconnection
from the Pulse
Generator
Fibrosis between
the nerve and
electrode
Electrode
detachment from
the nerve
Defective Pulse
Generator
Contact Cyberonics’ Clinical Technical
Support at
1 (866) 882-8804 (U.S. and
Canada) or +1 (281) 228-7330 (Worldwide).
Lead Impedance “HIGH”
Output Current “LOW”
Possible Causes:
Lead discontinuity
Lead disconnection
from the Pulse
Generator
Fibrosis between
the nerve and
electrode
Electrode
detachment from
the nerve
Defective Pulse
Generator
Lead Impedance “OK”
Output Current “LOW”
Possible Causes:
Increased
impedance in the
system
Lead Impedance “LOW”
Output Current “OK”
Possible Causes:
Short-circuit
condition within the
Lead
Defective Pulse
Generator
Error message encountered:
Lead Impedance “HIGH”
”A HIGH Lead Impedance has
been detected...”
Error message encountered:
Output Current “LOW”
”Programmed current is possibly
not being delivered...”
Lead Impedance “OK”
Output Current “OK”
Pulse Generator is delivering
stimulation as intended.