76 of 92 — 26-0008-1100/5
(Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
Figure 43.
Low Battery/End of Service Indications Prior to
Surgery, Outside of the Sterile Field - Initial
Implant or Pulse Generator Replacement
Battery Status Indicator or Message encountered:
IFI, N EOS, or EOS = Yes
.
Warning message - "The Intensified Follow-up Indicator (IFI) has been set for the Pulse Generator. More
frequent clinical monitoring is recommended. If you have any questions, please consult the Physician's Manual
or contact Cyberonics."
Warning message - "The Pulse Generator is nearing End of Service (N EOS). It is recommended that the Pulse
Generator be replaced as soon as possible. If you have any questions, please consult the Physician's Manual or
contact Cyberonics."
Warning message - "The Pulse Generator has reached End of Service (EOS) and is NOT supplying stimulation.
It is recommended that the Pulse Generator be replaced immediately. If you have any questions, please consult
the Physician's Manual or contact Cyberonics."
Warning message – “The Pulse Generator is currently disabled due to a Vbat< EOS threshold. Note that the
generator is NOT supplying stimulation. It is recommended that you contact Cyberonics or refer to the
Physician’s Manual.”
Perform System Diagnostics or Generator
Diagnostics with device in sterile package.
Record results.
Evaluate Diagnostic Results
Lead Impedance “any”
Impedance Value “any”
IFI = No
Proceed with implant.
Contact Cyberonics’ Clinical Technical
Support at
1 (866) 882-8804 (U.S. and
Canada) or +1 (281) 228-7330 (Worldwide)
.
Wait approximately 30 minutes
with Generator at room
temperature.
First time...
Second time...
Battery Status Indicator or Message encountered:
IFI, N EOS, or EOS = Yes
.
Warning message - "The Intensified Follow-up Indicator (IFI) has been set for
the Pulse Generator. More frequent clinical monitoring is recommended. If you
have any questions, please consult the Physician's Manual or contact
Cyberonics."
Warning message - "The Pulse Generator is nearing End of Service (N EOS). It
is recommended that the Pulse Generator be replaced as soon as possible. If
you have any questions, please consult the Physician's Manual or contact
Cyberonics."
Warning message - "The Pulse Generator has reached End of Service (EOS)
and is NOT supplying stimulation. It is recommended that the Pulse Generator
be replaced immediately. If you have any questions, please consult the
Physician's Manual or contact Cyberonics."
Warning message – “The Pulse Generator is currently disabled due to a Vbat<
EOS threshold. Note that the generator is NOT supplying stimulation. It is
recommended that you contact Cyberonics or refer to the Physician’s Manual.”