26-0008-1100/5
(Worldwide)
— 15 of 92
Physician’s Manual — VNS Therapy Programming Software (8.1)
3. The
New
button, when tapped, displays the range of possible settings
for that parameter
Figure 3.
PARAMETER Screen
↑
Inactivity Timeout Indicator (Progress Bar)
Note:
The inactivity
timeout function is not
used in Models 103,
104, and 105.
For the Models 100, 101, 102 Pulse Generators, the progress bar across the
bottom of the screen is an “inactivity timeout” indicator. When the bar
reaches the right margin, the most recent interrogation will have expired
and the warning shown in Figure 4 on page 16 will be displayed. Also, it
will be necessary to reinterrogate the Pulse Generator before
programming. The inactivity timeout period can be adjusted on the USER
PREFERENCES Screen. See See “User Preferences Screen ” on page 54.
Models 100, 101, 102 Models 103, 104, 105