16 of 92 — 26-0008-1100/5
(Worldwide)
Physician’s Manual — VNS Therapy Programming Software (8.1)
Note:
For more
information on the
timeout function, see
“User Preferences
Screen ” on page 54.
Figure 4.
TIMEOUT WARNING Screen
Caution:
LivaNova
recommends that you
reinterrogate the Pulse
Generator after each
programming event to
verify that the
programming occurred
as planned.
Note that the timeout bar has advanced to the right margin.
In addition to the eight programmable parameters, the PARAMETER
screen will also indicate battery status for the Pulse Generator. The
Programming Software has one battery status indicator, Near End of
Service (Near EOS), for Models 100, 101, and 102 Pulse Generators. The
Programming Software has three indicators, Intensified Follow-up
Indicator (IFI), Near End of Service (N EOS), and End of Service (EOS),
for Models 103, 104, and 105 Pulse Generators. On the PARAMETER
screen, the Near EOS status for Models 100, 101, and 102 Pulse
Generators is displayed on the lower portion of the screen (see Figure 3 on
page 15). For Model 103, 104, and 105 Pulse Generators, the remaining
generator battery power is indicated by the battery icon located in the top
left portion of the PARAMETER screen (see Figure 3 on page 15). The
battery icon is shaded and colored to indicate various levels of remaining
generator battery power. As the battery level decreases and the Pulse
Generator enter states of IFI = Yes, N EOS = Yes or EOS = Yes, an
additional indicator is displayed at the bottom of the PARAMETER screen
(see Figure 5 on page 17).