Problem
Nurse Action
Effect/Next Steps
Patient feels that they are / nurse thinks
that the patient is sitting directly on the
bed frame and not being supported by
the mattress.
Use COMFORT ADJUSTMENT controls
to increase air pressure in the mattress
areas A & B (see the Intructions for Use).
Patient should now look/feel that they are
not in contact with the mattress support
platform.
If in doubt, this can be checked by mo
-
ving or asking the patient to move slightly
from side to side. They should move fairly
easily.
If this is not the case and there are no
mattress alarms, consider repositio
-
ning the patient or using an alternative
mattress that can support heavier weight
patients.
Patient‘s rest disturbed by mattress auto
-
matic re-optimising.
Use MOBILE Mode.
Ensure the patient feels more comforta-
bly.