9.5 Fault Codes (OptiCare X integrated mattress)
Fault Code
Type of the Fault
Immediate Action of Operator
Corrective Action of Operator
P2
Major Inflate Fault
(air leak)
Check both manual CPR valves for closing. If
problem is still present, transfer patient to ano-
ther surface.
If problem is still present,
contact service department
approved by manufacturer!
P3
Deflate Fault
Wait three minutes until fault self-clear.
If problem is still present, transfer patient to
another surface.
If problem is still present,
contact service department
approved by manufacturer!
P4
Pressure Sensor Fault
Transfer patient to another surface.
Contact service department
approved by manufacturer!
P5
Overpressure Fault
Transfer patient to another surface.
Contact service department
approved by manufacturer!
P6
BOD Valve Fault
Transfer patient to another surface.
Contact service department
approved by manufacturer!
P7
Blower Fault
No MCM function, assess the effect on patient.
Contact service department
approved by manufacturer!
P8
Clogged Blower Filter
No MCM function, assess the effect on patient.
Contact service department
approved by manufacturer!
P9
Communication Fault
Contact service department approved by manufacturer!
P12
Optimization Out of Range
Transfer patient to another surface.
Contact service department
approved by manufacturer!
P14
Overcurrent of Solenoid
Transfer patient to another surface.
Contact service department
approved by manufacturer!
P92
System Control Unit
Overheating
Transfer patient to another surface.
Contact service department
approved by manufacturer!
P94
Pressure Calibration Fault
Transfer patient to another surface.
Contact service department
approved by manufacturer!
P98
Solenoid Fail
Transfer patient to another surface.
Contact service department
approved by manufacturer!
P100
Find the corresponding graphic pop-up in the chapter „9.4.13 STATUSES (Multiboard X - LCD touchscreen)“ on
the page 46 and follow the instructions related to it.
P101
P102
P105
P106
P107
P108