9.4.11 AUTOMATIC PRESSURE ADJUSTMENT ERROR
Automatic mattress internal pressure adjustment in MOBILE Mode and PRONE Mode is not available when the Automatic Pressure
Adjustment Error occurs. It is necessary to adjust the mattress internal pressure manually when this error occurs.
To eliminate this error:
►
Check if the scales are stabilized and if the scales works properly.
►
Check if the taring was performed without patient on the mattress.
►
Contact the service department of the manufacturer if this error remains.
Fig. AUTOMATIC PRESSURE ADJUSTMENT ERROR