OptiCare X for Multicare X
9.4.9 CPR MODE (CPR ACTIVATED)
When CPR is activated, the mattress will deflate and chest compression can start immediately.
CPR Mode period is limited to 1 hour and this period is counted down on the CPR Mode Screen.
OptiCare X inflates automatically when this time is up.
It is necessary to wait for the mattress inflation after switching from CPR Mode to OPTIMIZE Mode.
To deactivate CPR Mode:
►
press CPR CANCEL Icon.
The mattress will inflate again and return to the mode it was in before CPR started or to the mode it was in before MAX Mode if the
mattress was in the MAX Mode before the CPR Mode.
Fig. CPR MODE
Fig.
Pop-up meaning mattress inflation after switching from CPR Mode to OPTIMIZE Mode
25. 12. 25
22:58
CPR
CANCEL
CPR ACTIVATED
THE MATTRESS IS DEFLATED
AUTO INFLATE IN: 1:00:00
MOBILE
PRONE
MAX
OPTIMIZE
Do not use mattress
until fully inflated.
MATTRESS INFLATION